![Pfeiffer Vacuum Prisma 80 QMS 200 Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/pfeiffer-vacuum/prisma-80-qms-200/prisma-80-qms-200_operating-instructions-manual_1534984006.webp)
6
BG 805 204 BE / B (0203) QMS200.oi
2 Technical Data
Safety
QME/QMA
EN 61010-1:1993
IEC1010-1:1990 +A1:1992
Protection class I
SP 200
DIN EN 60950
IEC 950, UL 1950
CSA 22.2 No. 234
EMC
Immunity
Emission
EN 50082-2
EN 50081-1
Operation
Indoors only
Altitude max.
2000 m NN
Degree of protection
IP 30
Pollution degree
2
Temperatures
Storage/transport
–25 … +70 °C
Operation
QME
SP
0 … +40 °C
0 … +40 °C
Relative humidity
Max. 80% up to 31 °C,
Linearly decreasing to 50% at 40 °C
Mass range
1 … 80 u
Detector type
Faraday
Detection limit
<1
×
10
-12
mbar
Unit resolution
(valley, full mass range)
at 10% peak height
adjustable to
<10%
0.5 … 2.5 u
Sensitivity to argon
>1
×
10
-3
A/mbar
Contribution to
neighboring mass 40/41
<10 ppm
Operating pressure
≤
1
×
10
-4
mbar
≤
1
×
10
-3
mbar at an emission <200
µ
A
Reproducibility of peak ratio
(constant conditions over 8 h,
N
2
and Ar from air)
±0.5%
QMA operating temperature
0 … +150 °C
Bakeout temperature
(QME removed)
≤
200 °C
HS ion source
Open design, radial symmetry,
2 filaments (yttrium oxide), degassing by
electron bombardment (10 mA, 300 V)
Mounting orientation
Any (
→
"
12)
Weight
3 kg
Connection flange
DN 40 CF
2.1 Standards Complied
With
2.2 Ambient Conditions
2.3 Prisma™ 80
Содержание Prisma 80 QMS 200
Страница 34: ...34 BG 805 204 BE B 0203 QMS200 oi Notes ...