BG 805 204 BE / B (0203) QMS200.oi
DANGER
The analyzer flange must be properly grounded.
•
Connect it with all 6 screws to the properly grounded vacuum system.
•
If the vacuum system is not grounded, connect the QMA flange (grounding
screw M4) to ground by means of a separate ground conductor.
Establish this connection with yellow/green insulated or bare stranded
copper wire:
•
2.5 mm
²
if mechanically protected according to DIN VDE 110 T540
•
4.0 mm
²
if not protected
Caution
The QMA should preferably be mounted horizontally (with the guiding groove
downward). This results in optimum protection of the QME against falling objects
and allows easy installation and excellent access to the front panel.
Caution
Caution: dirt sensitive area
Touching the product or parts thereof with one's bare hands increases
the desorption rate.
Always wear clean, lint-free gloves and use clean tools when working
in this area.
Transport tube
Mounting tool
!
Set the QMA into the sup-
plied mounting tool.
"
Carefully remove the trans-
port tube from the QMA and
save it for possible future
use.
#
Flange the analyzer to the
vacuum chamber.
3.2.3 Installing the QMA
Removing the transport tube
and flanging the analyzer to
the vacuum chamber