BG 805 204 BE / B (0203) QMS200.oi
13
DANGER
The QME may be mounted only to a correctly installed QMA as the only ground
connection is established via the QMA flange.
The 24 V feeder line from the SP 200 must never be connected if the QMA is not
screw fastened to the QME.
Guide groove
Grounding tab
!
With a 4 mm hexagon socket screw key loosen the two screws on the black
plastic connection piece of the QME so that the moving part has a play of
2 … 3 mm.
"
Carefully slide the QME onto the mounted QMA up to the notch mark. Make
sure the positioning is correct (watch the grounding tab of the QME and the
guide groove of the QMA). Do not apply excessive force.
#
Firmly tighten the two hexagon socket screws. These screws secure the
QME mechanically and ensure electrical safety by providing ground contact.
WARNING
All electrical connections must have cable strain relief. Route the cables sepa-
rately from EMI sources.
WARNING
For reasons of electromagnetic compatibility, a single central grounding point is
strongly recommended for all connected units (pumping station, SP 200, com-
puter, recorders, etc.). A multiple AC power outlet or a common line voltage dis-
tributor with power switch fulfills this requirement in a simple manner.
3.2.4 Installing the QME
3.3 Electrical Connections
3.3.1 Connecting the QME