7.3.1 Dismantling connections
Safely disconnect the device from the mains before performing any work
1. Switch the equipment off.
2. Wait until all components have come to a complete standstill.
3. Switch off the main switch.
4. Disconnect the mains plug from the equipment.
1
2
3
4
5
Fig. 13:
Removing HiCube Eco connections
1
Mains cable
4
Connecting cable connector
2
Main switch
5
Accessories control cable
3
Multi-functional connection "X3"
1. Remove the connector from the multi-functional connection on the electronic drive unit of the tur-
bopump.
2. Unfasten and remove the accessories control cables from the connector.
7.3.2 Removing the turbopump
4
3
2
5
1
Fig. 14:
HiCube Eco | Dismantling the turbopump
1
Mounting plate
4
Fore-vacuum connection
2
Turbopump, complete
5
Fore-vacuum line
3
Retaining bolt with washer
1. Unfasten and remove the fore-vacuum line from the turbopump.
2. Ensure that the fore-vacuum hose is not bent or damaged.
3. Unscrew all retaining bolts (x4) and washers from the mounting plate.
Maintenance