9
Product description
4.3 Range of application
The pumping station HiCube Classic must be installed and operated in the follow-
ing ambient conditions.
Installation location
weather protected (indoors)
Temperature
+5 °C to +40 °C (up to +35 °C with air cooling)
Protection category
IP 20
Protection class
I
Relative humidity
max. 80 %, at T
≤
31 °C, up to max. 50% at T
≤
40 °C
Atmospheric pressure:
77 kPa - 106 kPa
Installation altitude
2000 m max.
Degree of pollution
2
Overvoltage category
II
Connection voltage TC
24 or 48 VDC ±5%
Limit value for
HiPace 80
HiPace 300
HiPace 400
HiPace 700
Permissible magnetic field
max.
3.3 mT
5.5 mT
6 mT
6 mT
Permissible irradiated heat
input
3 W
8 W
14 W
14 W