Section G: I/O Terminal Block Description 

 

12 

G:

 

I/O

 

T

ERMINAL 

B

LOCK 

D

ESCRIPTION

 

The I/O terminal blocks are found on the bottom panel of the cooler: 
 

TB1

  TB1 is the standard analog output (low voltage DC output) for all Classic 

Series Thermo-Electric Coolers.   Model M225 has two active 5” heat exchangers.  
The output is 0vdc to 2.5vdc for a temperature range of 0

°

C to 25

°

C.   

C

 

Terminal 1 is earth ground.  This terminal should be used to ground the shield of the 
shielded twisted pair cable that is used to connect the analog output to a receiving 
instrument. 

C

 

Terminal 2 is the signal return. 

C

 

Terminal 3 is the channel 1 output. 

C

 

Terminal 4 is the channel 2 output. 

 

TB2

  TB2 is the I/O terminal block (low level DC voltage or current or no voltage 

contact) used for all installed options, such as the alarm relay/water slip option or 
auxiliary analog output option. 

C

 

Terminal 1 is earth ground.  This terminal should be used to ground the shield of the 
shielded twisted pair cable that is used to connect the analog output board of the 
auxiliary analog output to a receiving instrument. 

C

 

Terminal 2 is the signal return for the auxiliary analog voltage output. 

C

 

Terminal 3 is the auxiliary analog voltage output. 

C

 

Terminal 4 is the negative side of the auxiliary analog 4-20mA output. 

C

 

Terminal 5 is the positive side of the auxiliary analog 4-20mA output. 

C

 

Terminals 6 & 7 are the dry contact form A relay output used for computer sense for 
the alarm relay/water slip option. 

C

 

Terminals 8 & 9 are the water slip sensor input for the alarm relay / water slip 
(moisture carryover) option.   

C

 

Terminal 10 is earth ground.  This terminal should be used to ground the shield of the 
twisted pair cable that is used to connect the water slip sensor to terminals 8 & 9. 

 

TB3

  TB3 is the switched AC terminal block (line AC voltage output) used for 

sample pump control when the alarm relay/water slip option is installed. 

C

 

Terminal 1 is the AC LINE voltage. 

C

 

Terminal 2 is the AC NEUTRAL. 

C

 

Terminal 3 is the EARTH GROUND. 

 

TB4

 

Note:

 The description for TB4 is the same as for TB2.  TB4 is used for 

Channel 2's I/O. 
 

TB5

 

Note:

 The description for TB5 is the same as for TB3.  TB5 is used for 

Channel 2's switched AC.

Содержание HORIBA Baldwin M225HHD

Страница 1: ...0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com HORIBA INSTRUCTION MANUAL CLASSIC THERMO ELECTRIC COOLERS...

Страница 2: ...ple of Operation 6 D Installation 9 E Start up Procedure 10 F LED Indicators 11 G I O Terminal Block Description 12 H Test Adjustment Procedures 13 I Description of Options 16 J New Jersey Thermocoupl...

Страница 3: ...BTU Hr 118kJ Hr Dimensions 11 20 x 7 25 x 11 20 in HWD 28 5 x 18 4 x 28 5 cm Weight 17 lbs 7 7 kg Maximum Inlet Sample Temperature 392 F 200 C SS Glass Impingers 280 F 138 C Kynar Impinger Maximum Inl...

Страница 4: ...rable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer shall conduct all testing and analysis necessary to val...

Страница 5: ...yzer performance specification Perma Pure coolers provide this constant low water concentration resulting in an accurate component gas measurement All Baldwin Series coolers use thermo electric elemen...

Страница 6: ...changer simply by slipping it out of the cooling block by hand The heat transfer block cools the heat exchanger through the heat pumping action of the peltier element The heat transfer block is on the...

Страница 7: ...r through the bottom drain connection Particulate matter passing through the sample cooler is removed by an optional Perma Pure pre filter located downstream from the cooler along with an optional wat...

Страница 8: ...tings are not used The inlet and outlet tubing for all metal or Kynar heat exchangers is 1 4 NPT the user should always use the compression type fittings provided for that purpose by the factory The i...

Страница 9: ...achieved After approximately 3 minutes the set point of 5 C 41 F is achieved The sample gas flow may be started immediately after the READY Green LED comes on The Model M225 is virtually maintenance f...

Страница 10: ...d on by other devices When the impinger temperature is below 10 C 50 F the ready LED s will be on Above this temperature the ready LED s will be off The middle green LED s applicable when the alarm re...

Страница 11: ...onnect the analog output board of the auxiliary analog output to a receiving instrument C Terminal 2 is the signal return for the auxiliary analog voltage output C Terminal 3 is the auxiliary analog v...

Страница 12: ...Channel 1 thermocouple input TB1 terminal inserting the yellow wire into the terminal marked Y and the red wire into the terminal marked R Turn on the generator The red LED should turn off Connect a...

Страница 13: ...from TB5 and TB6 on the auxiliary analog output board The thermocouple generator should not be connected to either of the thermocouple inputs at this time Connect power to the cooler After 10 to 20 se...

Страница 14: ...er to 200ma range Connect the ammeter to TB1 the channel 1 current output terminal the black lead to the negative terminal and the red lead to the positive terminal 4a Adjust POT4 the zero pot to obta...

Страница 15: ...ed for computer sensing and is NOT intended for the controlling of electrical loads The third output is a 6 amp DPST form C dry contact relay This relay can be used for sample pump or other heavier el...

Страница 16: ...the NJ identifies the upgrade for a NJ type thermocouple The K identifies the thermocouple itself as a type K thermocouple The 5 is the height of the heat exchangers In addition Perma Pure offers an...

Страница 17: ...erature but is below ready temperature System loading Calibration and set temperature adjustment Ensure system loading is not exceeding cooler capacity Adjust as necessary Impinger temperature cycles...

Страница 18: ...For further service assistance contact Perma Pure LLC P O Box 2105 8 Executive Drive 08755 Toms River NJ 08754 Tel 800 337 3762 toll free U S Tel 732 244 0010 Fax 732 244 8140 Email info permapure com...

Страница 19: ...nger 5 Durinert 3CXG 005 Heat Exchanger 5 Glass threaded w fittings 3CXK 001 Heat Exchanger 5 Kynar 3CXS 001 Heat Exchanger 5 Stainless Steel 3KPE 004 Peltier Element Kit 40 mm 1PSD 027 Power Supply 2...

Страница 20: ...21 APPENDIX A CLASSIC MODEL M225HHD...

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ...2 2M C2 1uf Q6 2N7000 VCC R1 2 2M C1 1uf Q3 2N7000 VCC Q1 2N7000 Q2 2N7000 Q4 2N7000 Q5 2N7000 CHN1_READY TC_1_OK CHN2_READY TC_2_OK M class Relay Board 1 3 5 7 RELAY2 1 3 5 7 RELAY3 1 2 TB4 1 2 TB1...

Страница 26: ...Perma Pure LLC Tel 732 244 0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com...

Отзывы: