background image

 

Section C: Principle of Operation 

 

       6 

C:

 

P

RINCIPLE OF 

O

PERATION

 

Thank you for purchasing the Perma Pure Baldwin™ Classic Series Thermo-Electric 
Cooler.  Our Classic Model M225 features a unique slim design leaving additional 
space to install or access other sample conditioning system components.  A unique 
drop-down door on the M225 provides easy access to electronic boards and the 
power supply.  The electronic control board is mounted on the door for easy access. 
 
The process of sampling combustion product stack gas or exhaust from internal 
combustion engines requires a method to remove the moisture from the sample, 
without removing the gas components of interest.  The Baldwin-Series Classic 
Thermo-Electric Cooler is an ideal way to decrease the dew point of combustion 
gases to a repeatable, stable, constant low dewpoint.  This cooler prevents water 
condensation in sample pre-filters, sample pumps, and gas analyzers.  For gas 
analyzers where water vapor is an interferent, a stable, repeatable dewpoint 
becomes a part of the gas analyzer performance specification.  Perma Pure coolers 
provide this constant low water concentration, resulting in an accurate component 
gas measurement. 
 
All Baldwin-Series  coolers use thermo-electric elements (Peltiers) to cool the 
sample gas to the desired dew point temperature.  A Classic Thermo-Electric Cooler 
is best illustrated as a small 
heat pump with no moving 
parts.  The Peltiers operate on 
direct current and may be used 
for heating or cooling by 
reversing the direction of 
current flow.  This is achieved 
by moving heat from one side 
of the module to the other with 
current flow and the laws of 
thermodynamics.  A typical 
single stage Peltier (Figure 1) 
consists of two ceramic plates 
with p- and n-type 
semiconductor material 
(bismuth telluride) between the 
plates.  The elements of 
semiconductor material are 
connected electrically in series and thermally in parallel. 

 

When a positive DC voltage is applied to the n-type thermo-electric element, 
electrons pass from the p- to the n-type thermo-electric element and the cold side 
temperature will decrease as heat is absorbed.  The heat absorption (cooling) is 
proportional to the current and the number of thermo-electric elements.  This heat is 

Figure 1: Thermo-electric element (Peltier)

Содержание HORIBA Baldwin M225HHD

Страница 1: ...0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com HORIBA INSTRUCTION MANUAL CLASSIC THERMO ELECTRIC COOLERS...

Страница 2: ...ple of Operation 6 D Installation 9 E Start up Procedure 10 F LED Indicators 11 G I O Terminal Block Description 12 H Test Adjustment Procedures 13 I Description of Options 16 J New Jersey Thermocoupl...

Страница 3: ...BTU Hr 118kJ Hr Dimensions 11 20 x 7 25 x 11 20 in HWD 28 5 x 18 4 x 28 5 cm Weight 17 lbs 7 7 kg Maximum Inlet Sample Temperature 392 F 200 C SS Glass Impingers 280 F 138 C Kynar Impinger Maximum Inl...

Страница 4: ...rable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer shall conduct all testing and analysis necessary to val...

Страница 5: ...yzer performance specification Perma Pure coolers provide this constant low water concentration resulting in an accurate component gas measurement All Baldwin Series coolers use thermo electric elemen...

Страница 6: ...changer simply by slipping it out of the cooling block by hand The heat transfer block cools the heat exchanger through the heat pumping action of the peltier element The heat transfer block is on the...

Страница 7: ...r through the bottom drain connection Particulate matter passing through the sample cooler is removed by an optional Perma Pure pre filter located downstream from the cooler along with an optional wat...

Страница 8: ...tings are not used The inlet and outlet tubing for all metal or Kynar heat exchangers is 1 4 NPT the user should always use the compression type fittings provided for that purpose by the factory The i...

Страница 9: ...achieved After approximately 3 minutes the set point of 5 C 41 F is achieved The sample gas flow may be started immediately after the READY Green LED comes on The Model M225 is virtually maintenance f...

Страница 10: ...d on by other devices When the impinger temperature is below 10 C 50 F the ready LED s will be on Above this temperature the ready LED s will be off The middle green LED s applicable when the alarm re...

Страница 11: ...onnect the analog output board of the auxiliary analog output to a receiving instrument C Terminal 2 is the signal return for the auxiliary analog voltage output C Terminal 3 is the auxiliary analog v...

Страница 12: ...Channel 1 thermocouple input TB1 terminal inserting the yellow wire into the terminal marked Y and the red wire into the terminal marked R Turn on the generator The red LED should turn off Connect a...

Страница 13: ...from TB5 and TB6 on the auxiliary analog output board The thermocouple generator should not be connected to either of the thermocouple inputs at this time Connect power to the cooler After 10 to 20 se...

Страница 14: ...er to 200ma range Connect the ammeter to TB1 the channel 1 current output terminal the black lead to the negative terminal and the red lead to the positive terminal 4a Adjust POT4 the zero pot to obta...

Страница 15: ...ed for computer sensing and is NOT intended for the controlling of electrical loads The third output is a 6 amp DPST form C dry contact relay This relay can be used for sample pump or other heavier el...

Страница 16: ...the NJ identifies the upgrade for a NJ type thermocouple The K identifies the thermocouple itself as a type K thermocouple The 5 is the height of the heat exchangers In addition Perma Pure offers an...

Страница 17: ...erature but is below ready temperature System loading Calibration and set temperature adjustment Ensure system loading is not exceeding cooler capacity Adjust as necessary Impinger temperature cycles...

Страница 18: ...For further service assistance contact Perma Pure LLC P O Box 2105 8 Executive Drive 08755 Toms River NJ 08754 Tel 800 337 3762 toll free U S Tel 732 244 0010 Fax 732 244 8140 Email info permapure com...

Страница 19: ...nger 5 Durinert 3CXG 005 Heat Exchanger 5 Glass threaded w fittings 3CXK 001 Heat Exchanger 5 Kynar 3CXS 001 Heat Exchanger 5 Stainless Steel 3KPE 004 Peltier Element Kit 40 mm 1PSD 027 Power Supply 2...

Страница 20: ...21 APPENDIX A CLASSIC MODEL M225HHD...

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ...2 2M C2 1uf Q6 2N7000 VCC R1 2 2M C1 1uf Q3 2N7000 VCC Q1 2N7000 Q2 2N7000 Q4 2N7000 Q5 2N7000 CHN1_READY TC_1_OK CHN2_READY TC_2_OK M class Relay Board 1 3 5 7 RELAY2 1 3 5 7 RELAY3 1 2 TB4 1 2 TB1...

Страница 26: ...Perma Pure LLC Tel 732 244 0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com...

Отзывы: