Section H: Test & Adjustment Procedures 

 

14 

B. Display Board (not standard on the Model M225)

 

 
(1)  

WARNING

: Before connecting power to the cooler, be aware of the 

HAZARDOUS LIVE VOLTAGE on the control board.  Disconnect power from the 
cooler before opening the swing-down "L-door".  Remove display cover from the 
front of the swing-down "L-door".  Connect a voltmeter between the ground test 
point, TP7 on the main control board and TP1 on the display board, the red lead to 
TP1 and the black lead to TP7. 
 
(2)  Connect power to the cooler and allow the temperature of the cooler to stabilize 
around 5

°

C. 

 
(3)  Adjust POT1 on the display board so the display corresponds to the voltmeter 
reading.  That is, if the voltmeter reads 0.5vdc, adjust the pot so that the display 
reads 0.5. 
 

C. Auxiliary Analog Output Board (not standard on the model M225)

 

 
(1)

  WARNING

: Before connecting power to the cooler, be aware of the 

HAZARDOUS LIVE VOLTAGE on the control board.  Disconnect power from the 
cooler before opening the swing-down "L-door".  Remove the thermocouples from 
TB5 and TB6 on the auxiliary analog output board.  The thermocouple generator 
should not be connected to either of the thermocouple inputs at this time.  Connect 
power to the cooler. After 10 to 20 seconds, the red LED(s) (thermocouple failure 
indicator) on the auxiliary analog output board should be on. 
 
(2)  Connect the thermocouple generator to the Channel 1 thermocouple input (TB5 
terminal), inserting the yellow wire into the terminal marked Y and the red wire into 
the terminal marked R.  Turn on the generator.  The red LED (CR5) should turn off.  
Set the thermocouple generator to 0

°

C. 

 
For Voltage Output 
(3)  If the auxiliary analog output board is set for voltage output, set the voltmeter to 
20vdc range.  Connect the voltmeter to TB2 (the channel 1 voltage output terminal), 
the black lead to the negative (-) terminal and the red lead to the positive (+) 
terminal. 
 
(4)  Adjust POT4 (the zero pot) to obtain a 0vdc reading on the voltmeter. 
 
(5)  Set the thermocouple generator to 25

°

C. 

 
(6)  Adjust POT1 (the span pot) to obtain a full scale voltage reading on the 
voltmeter.  This value will depend on the gain that was specified for the cooler at the 
time of manufacture.  Available gains are 0-1.0vdc, 0-2.5vdc, 0-10.0vdc for a 0

°

C-

25

°

C temperature range. 

 

Содержание HORIBA Baldwin M225HHD

Страница 1: ...0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com HORIBA INSTRUCTION MANUAL CLASSIC THERMO ELECTRIC COOLERS...

Страница 2: ...ple of Operation 6 D Installation 9 E Start up Procedure 10 F LED Indicators 11 G I O Terminal Block Description 12 H Test Adjustment Procedures 13 I Description of Options 16 J New Jersey Thermocoupl...

Страница 3: ...BTU Hr 118kJ Hr Dimensions 11 20 x 7 25 x 11 20 in HWD 28 5 x 18 4 x 28 5 cm Weight 17 lbs 7 7 kg Maximum Inlet Sample Temperature 392 F 200 C SS Glass Impingers 280 F 138 C Kynar Impinger Maximum Inl...

Страница 4: ...rable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer shall conduct all testing and analysis necessary to val...

Страница 5: ...yzer performance specification Perma Pure coolers provide this constant low water concentration resulting in an accurate component gas measurement All Baldwin Series coolers use thermo electric elemen...

Страница 6: ...changer simply by slipping it out of the cooling block by hand The heat transfer block cools the heat exchanger through the heat pumping action of the peltier element The heat transfer block is on the...

Страница 7: ...r through the bottom drain connection Particulate matter passing through the sample cooler is removed by an optional Perma Pure pre filter located downstream from the cooler along with an optional wat...

Страница 8: ...tings are not used The inlet and outlet tubing for all metal or Kynar heat exchangers is 1 4 NPT the user should always use the compression type fittings provided for that purpose by the factory The i...

Страница 9: ...achieved After approximately 3 minutes the set point of 5 C 41 F is achieved The sample gas flow may be started immediately after the READY Green LED comes on The Model M225 is virtually maintenance f...

Страница 10: ...d on by other devices When the impinger temperature is below 10 C 50 F the ready LED s will be on Above this temperature the ready LED s will be off The middle green LED s applicable when the alarm re...

Страница 11: ...onnect the analog output board of the auxiliary analog output to a receiving instrument C Terminal 2 is the signal return for the auxiliary analog voltage output C Terminal 3 is the auxiliary analog v...

Страница 12: ...Channel 1 thermocouple input TB1 terminal inserting the yellow wire into the terminal marked Y and the red wire into the terminal marked R Turn on the generator The red LED should turn off Connect a...

Страница 13: ...from TB5 and TB6 on the auxiliary analog output board The thermocouple generator should not be connected to either of the thermocouple inputs at this time Connect power to the cooler After 10 to 20 se...

Страница 14: ...er to 200ma range Connect the ammeter to TB1 the channel 1 current output terminal the black lead to the negative terminal and the red lead to the positive terminal 4a Adjust POT4 the zero pot to obta...

Страница 15: ...ed for computer sensing and is NOT intended for the controlling of electrical loads The third output is a 6 amp DPST form C dry contact relay This relay can be used for sample pump or other heavier el...

Страница 16: ...the NJ identifies the upgrade for a NJ type thermocouple The K identifies the thermocouple itself as a type K thermocouple The 5 is the height of the heat exchangers In addition Perma Pure offers an...

Страница 17: ...erature but is below ready temperature System loading Calibration and set temperature adjustment Ensure system loading is not exceeding cooler capacity Adjust as necessary Impinger temperature cycles...

Страница 18: ...For further service assistance contact Perma Pure LLC P O Box 2105 8 Executive Drive 08755 Toms River NJ 08754 Tel 800 337 3762 toll free U S Tel 732 244 0010 Fax 732 244 8140 Email info permapure com...

Страница 19: ...nger 5 Durinert 3CXG 005 Heat Exchanger 5 Glass threaded w fittings 3CXK 001 Heat Exchanger 5 Kynar 3CXS 001 Heat Exchanger 5 Stainless Steel 3KPE 004 Peltier Element Kit 40 mm 1PSD 027 Power Supply 2...

Страница 20: ...21 APPENDIX A CLASSIC MODEL M225HHD...

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ...2 2M C2 1uf Q6 2N7000 VCC R1 2 2M C1 1uf Q3 2N7000 VCC Q1 2N7000 Q2 2N7000 Q4 2N7000 Q5 2N7000 CHN1_READY TC_1_OK CHN2_READY TC_2_OK M class Relay Board 1 3 5 7 RELAY2 1 3 5 7 RELAY3 1 2 TB4 1 2 TB1...

Страница 26: ...Perma Pure LLC Tel 732 244 0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com...

Отзывы: