background image

 

Section A: Specifications 

 

A:

 

S

PECIFICATIONS

 

 

 

 

Physical Description 

Single (Series) / Dual (Parallel) Channel System 
2 x 5” Heat Exchangers connected in series or parallel 
2 Active (cooled to 4

°

C) Heat Exchangers 

LCD temperature display (optional) 
 

Operating Specifications 

Sample Gas Flow Range 

3-5 LPM 

6.5-10.8 SCFH 

Inlet Dew Point at Rated 
Flow 

140

°

F @ 20% H

2

O, 5 LPM 

60

°

Maximum Cooling Rate 

112 BTU/Hr 

118kJ/Hr 

Dimensions 

11.20 x 7.25 x 11.20 in. HWD 

28.5 x 18.4 x 28.5 cm 

Weight 17 

lbs 

7.7 kg 

Maximum Inlet Sample 
Temperature 

392

°

(200

°

C)

 SS, Glass 

Impingers 
280

°

(138

°

C)

 Kynar Impinger 

Maximum Inlet Pressure 

45 psig 

3 bar / 2250 mmHg 

Maximum Heat Exchanger 
Pressure Drop 

<+1 in. H

2

Ambient Temperature 
Range 

33-104

°

0.56-40

°

Outlet Sample Gas Dew 
Point 

41

°

5

°

Inlet Tubing Connection 

 in. FPT  

Outlet Tubing Connection 

¼ in. FPT 

Drain Tubing Connection 

 in. FPT 

Voltage 

110 (220 optional) VAC 
50/60 Hz 

Thermoelectric Elements 

40 mm 

Power Supply 

250W 

Cooling Down Time 

Less than 3 minutes 

 
 
 
 
 

Содержание HORIBA Baldwin M225HHD

Страница 1: ...0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com HORIBA INSTRUCTION MANUAL CLASSIC THERMO ELECTRIC COOLERS...

Страница 2: ...ple of Operation 6 D Installation 9 E Start up Procedure 10 F LED Indicators 11 G I O Terminal Block Description 12 H Test Adjustment Procedures 13 I Description of Options 16 J New Jersey Thermocoupl...

Страница 3: ...BTU Hr 118kJ Hr Dimensions 11 20 x 7 25 x 11 20 in HWD 28 5 x 18 4 x 28 5 cm Weight 17 lbs 7 7 kg Maximum Inlet Sample Temperature 392 F 200 C SS Glass Impingers 280 F 138 C Kynar Impinger Maximum Inl...

Страница 4: ...rable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer shall conduct all testing and analysis necessary to val...

Страница 5: ...yzer performance specification Perma Pure coolers provide this constant low water concentration resulting in an accurate component gas measurement All Baldwin Series coolers use thermo electric elemen...

Страница 6: ...changer simply by slipping it out of the cooling block by hand The heat transfer block cools the heat exchanger through the heat pumping action of the peltier element The heat transfer block is on the...

Страница 7: ...r through the bottom drain connection Particulate matter passing through the sample cooler is removed by an optional Perma Pure pre filter located downstream from the cooler along with an optional wat...

Страница 8: ...tings are not used The inlet and outlet tubing for all metal or Kynar heat exchangers is 1 4 NPT the user should always use the compression type fittings provided for that purpose by the factory The i...

Страница 9: ...achieved After approximately 3 minutes the set point of 5 C 41 F is achieved The sample gas flow may be started immediately after the READY Green LED comes on The Model M225 is virtually maintenance f...

Страница 10: ...d on by other devices When the impinger temperature is below 10 C 50 F the ready LED s will be on Above this temperature the ready LED s will be off The middle green LED s applicable when the alarm re...

Страница 11: ...onnect the analog output board of the auxiliary analog output to a receiving instrument C Terminal 2 is the signal return for the auxiliary analog voltage output C Terminal 3 is the auxiliary analog v...

Страница 12: ...Channel 1 thermocouple input TB1 terminal inserting the yellow wire into the terminal marked Y and the red wire into the terminal marked R Turn on the generator The red LED should turn off Connect a...

Страница 13: ...from TB5 and TB6 on the auxiliary analog output board The thermocouple generator should not be connected to either of the thermocouple inputs at this time Connect power to the cooler After 10 to 20 se...

Страница 14: ...er to 200ma range Connect the ammeter to TB1 the channel 1 current output terminal the black lead to the negative terminal and the red lead to the positive terminal 4a Adjust POT4 the zero pot to obta...

Страница 15: ...ed for computer sensing and is NOT intended for the controlling of electrical loads The third output is a 6 amp DPST form C dry contact relay This relay can be used for sample pump or other heavier el...

Страница 16: ...the NJ identifies the upgrade for a NJ type thermocouple The K identifies the thermocouple itself as a type K thermocouple The 5 is the height of the heat exchangers In addition Perma Pure offers an...

Страница 17: ...erature but is below ready temperature System loading Calibration and set temperature adjustment Ensure system loading is not exceeding cooler capacity Adjust as necessary Impinger temperature cycles...

Страница 18: ...For further service assistance contact Perma Pure LLC P O Box 2105 8 Executive Drive 08755 Toms River NJ 08754 Tel 800 337 3762 toll free U S Tel 732 244 0010 Fax 732 244 8140 Email info permapure com...

Страница 19: ...nger 5 Durinert 3CXG 005 Heat Exchanger 5 Glass threaded w fittings 3CXK 001 Heat Exchanger 5 Kynar 3CXS 001 Heat Exchanger 5 Stainless Steel 3KPE 004 Peltier Element Kit 40 mm 1PSD 027 Power Supply 2...

Страница 20: ...21 APPENDIX A CLASSIC MODEL M225HHD...

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ...2 2M C2 1uf Q6 2N7000 VCC R1 2 2M C1 1uf Q3 2N7000 VCC Q1 2N7000 Q2 2N7000 Q4 2N7000 Q5 2N7000 CHN1_READY TC_1_OK CHN2_READY TC_2_OK M class Relay Board 1 3 5 7 RELAY2 1 3 5 7 RELAY3 1 2 TB4 1 2 TB1...

Страница 26: ...Perma Pure LLC Tel 732 244 0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com...

Отзывы: