background image

 

Section J: Troubleshooting 

 

18 

K:

 

T

ROUBLESHOOTING

 

 

Symptom Check 

Action 

No LED(s) and no fan. 

AC power input. 

Ensure that AC power is 
connected. 

No LED(s) and fan on. 

AC input fuse on control board. 
DC output fuse on control board. 
VC on control board. 

Replace fuse as necessary. 
Replace fuse as necessary. 
Replace control board. 
 

LED(s) on and no fan. 

AC input fuse on power supply. 
+12vdc TB4 on control board. 
 
 

Replace fuse as necessary. 
Replace power supply. 

Impinger remains at 
ambient temperature. 

Peltier current draw.  Should be 
above 6 amps. 

Replace Peltier element. 

Thermocouple failure LED is 
on. 

Thermocouple connection TB1, 2.  Ensure proper connection. 

Replace thermocouple. 

Impinger frozen and cooler 
indicates ambient 
temperature. 

Thermocouple placement in heat 
exchanger block. 

Ensure proper placement. 
Replace control board. 

Impinger does not reach set 
temperature, but is below 
ready temperature. 

System loading. 
 
Calibration and set temperature 
adjustment. 

Ensure system loading is not 
exceeding cooler capacity. 
Adjust as necessary. 

Impinger temperature cycles 
up and down. 

Peltier connections on control 
board. 

Ensure a firm connection on 
flag connectors on control 
board.  Ensure system loading 
is not exceeding cooler 
capacity. 

Ready LED does not come on 
when impinger is below 7°C. 

Ready temperature adjustment. 

Adjust as necessary. 

Water carryover in system. 

Impinger temperature.  Should be 
below 6°C. 

Ensure system loading is not 
exceeding cooler capacity. 

Slip LED does not come on 

(alarm relay/water slip option 
installed)

Water carryover in system. 
 
Water slip sensor connections. 

Ensure system loading is not 
exceeding cooler capacity. 
Ensure that all water slip sensor 
connections are made.   
Clean tip of sensor.   
Replace alarm relay/water slip 
board. 

Pump does not start.  Ready 
and slip LED(s) are on 

(alarm 

relay/water slip option 
installed)

Pump electrical connections. 

Ensure proper connections. 
Replace board. 

 
 
 
 
 
 

Содержание HORIBA Baldwin M225HHD

Страница 1: ...0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com HORIBA INSTRUCTION MANUAL CLASSIC THERMO ELECTRIC COOLERS...

Страница 2: ...ple of Operation 6 D Installation 9 E Start up Procedure 10 F LED Indicators 11 G I O Terminal Block Description 12 H Test Adjustment Procedures 13 I Description of Options 16 J New Jersey Thermocoupl...

Страница 3: ...BTU Hr 118kJ Hr Dimensions 11 20 x 7 25 x 11 20 in HWD 28 5 x 18 4 x 28 5 cm Weight 17 lbs 7 7 kg Maximum Inlet Sample Temperature 392 F 200 C SS Glass Impingers 280 F 138 C Kynar Impinger Maximum Inl...

Страница 4: ...rable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer shall conduct all testing and analysis necessary to val...

Страница 5: ...yzer performance specification Perma Pure coolers provide this constant low water concentration resulting in an accurate component gas measurement All Baldwin Series coolers use thermo electric elemen...

Страница 6: ...changer simply by slipping it out of the cooling block by hand The heat transfer block cools the heat exchanger through the heat pumping action of the peltier element The heat transfer block is on the...

Страница 7: ...r through the bottom drain connection Particulate matter passing through the sample cooler is removed by an optional Perma Pure pre filter located downstream from the cooler along with an optional wat...

Страница 8: ...tings are not used The inlet and outlet tubing for all metal or Kynar heat exchangers is 1 4 NPT the user should always use the compression type fittings provided for that purpose by the factory The i...

Страница 9: ...achieved After approximately 3 minutes the set point of 5 C 41 F is achieved The sample gas flow may be started immediately after the READY Green LED comes on The Model M225 is virtually maintenance f...

Страница 10: ...d on by other devices When the impinger temperature is below 10 C 50 F the ready LED s will be on Above this temperature the ready LED s will be off The middle green LED s applicable when the alarm re...

Страница 11: ...onnect the analog output board of the auxiliary analog output to a receiving instrument C Terminal 2 is the signal return for the auxiliary analog voltage output C Terminal 3 is the auxiliary analog v...

Страница 12: ...Channel 1 thermocouple input TB1 terminal inserting the yellow wire into the terminal marked Y and the red wire into the terminal marked R Turn on the generator The red LED should turn off Connect a...

Страница 13: ...from TB5 and TB6 on the auxiliary analog output board The thermocouple generator should not be connected to either of the thermocouple inputs at this time Connect power to the cooler After 10 to 20 se...

Страница 14: ...er to 200ma range Connect the ammeter to TB1 the channel 1 current output terminal the black lead to the negative terminal and the red lead to the positive terminal 4a Adjust POT4 the zero pot to obta...

Страница 15: ...ed for computer sensing and is NOT intended for the controlling of electrical loads The third output is a 6 amp DPST form C dry contact relay This relay can be used for sample pump or other heavier el...

Страница 16: ...the NJ identifies the upgrade for a NJ type thermocouple The K identifies the thermocouple itself as a type K thermocouple The 5 is the height of the heat exchangers In addition Perma Pure offers an...

Страница 17: ...erature but is below ready temperature System loading Calibration and set temperature adjustment Ensure system loading is not exceeding cooler capacity Adjust as necessary Impinger temperature cycles...

Страница 18: ...For further service assistance contact Perma Pure LLC P O Box 2105 8 Executive Drive 08755 Toms River NJ 08754 Tel 800 337 3762 toll free U S Tel 732 244 0010 Fax 732 244 8140 Email info permapure com...

Страница 19: ...nger 5 Durinert 3CXG 005 Heat Exchanger 5 Glass threaded w fittings 3CXK 001 Heat Exchanger 5 Kynar 3CXS 001 Heat Exchanger 5 Stainless Steel 3KPE 004 Peltier Element Kit 40 mm 1PSD 027 Power Supply 2...

Страница 20: ...21 APPENDIX A CLASSIC MODEL M225HHD...

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ...2 2M C2 1uf Q6 2N7000 VCC R1 2 2M C1 1uf Q3 2N7000 VCC Q1 2N7000 Q2 2N7000 Q4 2N7000 Q5 2N7000 CHN1_READY TC_1_OK CHN2_READY TC_2_OK M class Relay Board 1 3 5 7 RELAY2 1 3 5 7 RELAY3 1 2 TB4 1 2 TB1...

Страница 26: ...Perma Pure LLC Tel 732 244 0010 8 Executive Dr PO Box 2105 Tel 800 337 3762 toll free US Toms River NJ 08754 Fax 732 244 8140 www permapure com Email info permapure com...

Отзывы: