Using the RamanMicro 200 . 97
12.
Select Wavelength Calibration Verification, and then click OK.
The wavelength calibration verification routine begins. This will take a few minutes.
After completing the wavelength calibration verification routine, an audit log is
generated.
13.
Remove the polystyrene sample and store safely.
The instrument is now ready for use.
In the unlikely event that the wavelength calibration verification should fail, follow Figure 33.
Further Details on the wavelength calibration verification are provided as a spectral file
(*.sp), which is saved in
C:\pel_data\config\Raman\Calibration Files\Logs
folder.
SOP-12: Wavelength Calibration of the
RamanMicro 200
Wavelength calibration should only be performed if the instrument has failed a wavelength
calibration verification.
1.
Ensure the RamanMicro 200 is switched on.
If not follow SOP-1.
2.
Attach the external neon calibration accessory (L1320220) to the rear panel of the
spectrometer module.
Push the large connector into the socket labeled CALIBRATION or CAL ACCY and
carefully attach the small cable to the COL coupler below the LED FIBER 1 connector.
Figure 34 Attaching the calibration accessory
NOTE: As wavelength calibration is not required on a regular basis, the external neon
calibration accessory is not included in the basic RamanMicro 200 package. The
external neon calibration accessory is included in the RamanMicro Calibration Kit
(L1320266), which can be purchased separately.
NOTE: During the wavelength calibration, make sure that no light is going into the system.
Содержание Raman Micro 200 Series
Страница 1: ...RamanMicro 200 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Страница 5: ...Introduction...
Страница 10: ...10 RamanMicro 200 Series Getting Started Guide...
Страница 11: ...Safety Practices...
Страница 29: ...Raman History Theory and Instrumentation...
Страница 35: ...Preparing Your Facility...
Страница 41: ...Installing Your RamanMicro 200...
Страница 76: ...76 RamanMicro 200 Series Getting Started Guide...
Страница 77: ...Using the RamanMicro 200...
Страница 103: ...Maintenance and Troubleshooting...