82 .
RamanMicro 200 Series Getting Started Guide
If this glass fluorescence is a problem and it is not possible to use quartz containers, then it
is possible to acquire a spectrum of the container, and spectrally subtract this from any
spectra containing this fluorescent signature.
Sampling Slurries/Highly absorbing species
Please refer to
Sampling Liquids
. For sample immersion care taken in focusing the probe
within the sleeve can result in a dramatic improvement in spectral quality.
Sampling Gases
Although Raman scattering from gas samples is very weak, the FOP can be used to measure
gas phase under long acquisition times.
WARNING
The standard FOP is not designed for immersion use. The FOP is
sealed against minor splashes. However, immersion in solvents will
cause serious damage. Ensure an immersion sleeve is in place before
immersing your FOP; see Figure 29. Please contact PerkinElmer if you
have any questions in relation to sampling using the FOP.