61
Performing a Digestion
When assembling the vessel, visually inspect the metallic rupture disc. If there is
evidence of corrosion or other visible damage to the rupture discs, they must be
replaced.
A small bulge in the center of the rupture disc after use is normal and will not affect
its function. If this bulge exceeds 2mm in height or displays thinning at the edges
or top, it must be replaced.
Make sure the metallic rupture disc is properly centered and seated on the TFM™
seal with its edge sitting uniformly in the groove near the edge of the seal. Expand
the lip of the TFM™ seal using the included forming tool and then place the seal
with the rupture disc onto the vessel body. At this point, the cap can be placed on
the vessel and the vessel with sample is now ready to be placed in the Titan MPS.
Figure 5-4.
Expanding the lip on the TFM™ seal
Caution
Vessel minimum fill amount!
In each pressure vessel, no less than 5ml of liquid should be employed so
that the temperature sensor can determine the sample temperature
precisely.
The microwave oven requires a minimum load of approximately 20 mL of
liquid. The 20 mL minimum should be evenly spread amongst at least 4
vessels for each run in order to achieve the required total load. If the fluid
amount employed is too small, the oven may be damaged within a matter
of minutes! An inadequate liquid load can also result in rotor overheating
and in the worst case, rotor damage.
Содержание TITAN MPS
Страница 1: ...TITAN MPS Customer Users Guide ATOMIC SPECTROSCOPY ...
Страница 2: ......
Страница 3: ...Titan MPS Users Guide ...
Страница 55: ...45 Power Cord Options Buck and Boost Transformer 0 5 KVA Japan and U S Canada Connections ...
Страница 98: ...88 Operation Figure 5 27 Sample Details review 1 8 Figure 5 28 Sample Details review 9 16 ...
Страница 104: ...94 Operation Sample Details Entering individual sample information ...
Страница 105: ...95 Digestion Results Digestion Results Reviewing the data from completed digestions ...
Страница 106: ...96 Operation Change Rename Datafile Locating and or renaming the datafile used to save digestion data ...
Страница 107: ...97 Create Method Create Method The process of generating a new method Once saved the method is permanently changed ...
Страница 108: ...98 Operation Edit Method The process of editing a current method Once saved the method is permanently changed ...
Страница 109: ...99 Data Management Data Management Options available via the Data button on the Home screen ...
Страница 120: ...110 Maintenance ...
Страница 122: ...112 Maintenance ...
Страница 132: ...122 Appendix Login Login 1 Select the login access desired ...
Страница 133: ...123 Index of Controller Screens Login Passcode 1 Enter the login passcode ...
Страница 150: ...140 Appendix Review Method Back Return to the Digestion Graph screen Next Go to the Review Sample Details 1 8 screen ...
Страница 156: ...146 Appendix ...
Страница 160: ......