Installation
Hot acid vapors may be released.
The supplied vent tubes must therefore always be properly connected and
attached to an exhaust system.
Make sure the PFA tube is securely attached so that it cannot pop off its
connection fitting in case of a percussive pressure burst as occurs when a
rupture disc activates.
Note
The laboratory is responsible for ensuring that the exhaust system can
appropriately vent the released gases. Make sure that all ventilation
procedures conform to applicable laws.
Note
Important Note for North American Users:
Voltage of mains supply has to be in the range of 230V ± 15V AC. Even
with a maximum load of approximately 2700 W the supply-voltage should
not drop below 215 V AC. Low voltage can result in low microwave power
delivery, excessive magnetron or inverter heating and erratic temperature
and pressure measurements.
!
Warning
Содержание TITAN MPS
Страница 1: ...TITAN MPS Customer Users Guide ATOMIC SPECTROSCOPY ...
Страница 2: ......
Страница 3: ...Titan MPS Users Guide ...
Страница 55: ...45 Power Cord Options Buck and Boost Transformer 0 5 KVA Japan and U S Canada Connections ...
Страница 98: ...88 Operation Figure 5 27 Sample Details review 1 8 Figure 5 28 Sample Details review 9 16 ...
Страница 104: ...94 Operation Sample Details Entering individual sample information ...
Страница 105: ...95 Digestion Results Digestion Results Reviewing the data from completed digestions ...
Страница 106: ...96 Operation Change Rename Datafile Locating and or renaming the datafile used to save digestion data ...
Страница 107: ...97 Create Method Create Method The process of generating a new method Once saved the method is permanently changed ...
Страница 108: ...98 Operation Edit Method The process of editing a current method Once saved the method is permanently changed ...
Страница 109: ...99 Data Management Data Management Options available via the Data button on the Home screen ...
Страница 120: ...110 Maintenance ...
Страница 122: ...112 Maintenance ...
Страница 132: ...122 Appendix Login Login 1 Select the login access desired ...
Страница 133: ...123 Index of Controller Screens Login Passcode 1 Enter the login passcode ...
Страница 150: ...140 Appendix Review Method Back Return to the Digestion Graph screen Next Go to the Review Sample Details 1 8 screen ...
Страница 156: ...146 Appendix ...
Страница 160: ......