116
Troubleshooting
Control Unit
Problem
Possible Cause(s)
Corrective Action
Pressure reading is inaccu-
rate or fluctuates signifi-
cantly.
Pressure calibration is not
accurate.
Note: Below 5 bar, some fluc-
tuation is normal.
DPC cap is contaminated.
DPC cap is damaged or needs
inspection and cleaning.
Inspect and clean the DPC
glass parts and polarization
film. (See Chapter 6, Mainte-
nance, DPC Cap Mainte-
nance.)
Replace piston o-rings.
Temperature begins with val-
ues >80° C during digestion
start
Defect on the temperature
sensor.
Highly exothermic reaction.
Contact your PerkinElmer
Service Representative.
After adding acids, wait for
the reaction to taper off
before closing the caps.
Signals at unoccupied posi-
tions
Wrong number of vessels
configured.
Check and set number of ves-
sels prior to start.
Program stops during the start
phase "Turntable Error"
Vessels put in incorrectly.
Decoder defective.
Motor defective.
Put in vessels correctly.
Contact your PerkinElmer
Service Representative.
Contact your PerkinElmer
Service Representative.
Содержание TITAN MPS
Страница 1: ...TITAN MPS Customer Users Guide ATOMIC SPECTROSCOPY ...
Страница 2: ......
Страница 3: ...Titan MPS Users Guide ...
Страница 55: ...45 Power Cord Options Buck and Boost Transformer 0 5 KVA Japan and U S Canada Connections ...
Страница 98: ...88 Operation Figure 5 27 Sample Details review 1 8 Figure 5 28 Sample Details review 9 16 ...
Страница 104: ...94 Operation Sample Details Entering individual sample information ...
Страница 105: ...95 Digestion Results Digestion Results Reviewing the data from completed digestions ...
Страница 106: ...96 Operation Change Rename Datafile Locating and or renaming the datafile used to save digestion data ...
Страница 107: ...97 Create Method Create Method The process of generating a new method Once saved the method is permanently changed ...
Страница 108: ...98 Operation Edit Method The process of editing a current method Once saved the method is permanently changed ...
Страница 109: ...99 Data Management Data Management Options available via the Data button on the Home screen ...
Страница 120: ...110 Maintenance ...
Страница 122: ...112 Maintenance ...
Страница 132: ...122 Appendix Login Login 1 Select the login access desired ...
Страница 133: ...123 Index of Controller Screens Login Passcode 1 Enter the login passcode ...
Страница 150: ...140 Appendix Review Method Back Return to the Digestion Graph screen Next Go to the Review Sample Details 1 8 screen ...
Страница 156: ...146 Appendix ...
Страница 160: ......