Problem
Possible Cause(s)
Corrective Action
Vessels leak.
TFM seal is not sealing effec-
tively on the digestion vessel.
Expand the lip of the TFM
seal using the TFM seal tools
If the TFM seal is damaged,
replace the seal.
Expand the lip of the TFM
seal using the TFM seal tools
If the TFM seal is damaged,
replace the seal see Chapter 6
Maintenance, Replacing TFM
Seals and Metal Rupture
Discs.
Corroded rupture disks.
Corroded rupture disks
Turntable function faulty.
Sample quantity too big.
Spontaneous reaction.
DTC Temperature measure-
ment inaccurate.
Replace burst disks more fre-
quently.
Check turntable function.
Reduce sample weight.
Reduce maximum tempera-
ture.
Contact your PerkinElmer
Service Representative.
DPC not preventing rupture
discs from venting.
Measured pressure is inaccu-
rate.
Inspect and clean the glass
prism and glass ring in the
pressure-gauge cap.
Perform a Pressure Calibra-
tion.
Содержание TITAN MPS
Страница 1: ...TITAN MPS Customer Users Guide ATOMIC SPECTROSCOPY ...
Страница 2: ......
Страница 3: ...Titan MPS Users Guide ...
Страница 55: ...45 Power Cord Options Buck and Boost Transformer 0 5 KVA Japan and U S Canada Connections ...
Страница 98: ...88 Operation Figure 5 27 Sample Details review 1 8 Figure 5 28 Sample Details review 9 16 ...
Страница 104: ...94 Operation Sample Details Entering individual sample information ...
Страница 105: ...95 Digestion Results Digestion Results Reviewing the data from completed digestions ...
Страница 106: ...96 Operation Change Rename Datafile Locating and or renaming the datafile used to save digestion data ...
Страница 107: ...97 Create Method Create Method The process of generating a new method Once saved the method is permanently changed ...
Страница 108: ...98 Operation Edit Method The process of editing a current method Once saved the method is permanently changed ...
Страница 109: ...99 Data Management Data Management Options available via the Data button on the Home screen ...
Страница 120: ...110 Maintenance ...
Страница 122: ...112 Maintenance ...
Страница 132: ...122 Appendix Login Login 1 Select the login access desired ...
Страница 133: ...123 Index of Controller Screens Login Passcode 1 Enter the login passcode ...
Страница 150: ...140 Appendix Review Method Back Return to the Digestion Graph screen Next Go to the Review Sample Details 1 8 screen ...
Страница 156: ...146 Appendix ...
Страница 160: ......