SGM700 Digitizer
Configuration Menu -Ind- -continiue-
The following screen is vissable:
In
Ind 3
you set the
Stable time
Indicator gives stable signal when weigher value
is stable within the range set in Ind 2 and time set with Ind 3.
Choose a value between: 000.000 – 16.959 seconds.
To change the value press key 2 < 2 sec.
The following screen is vissable:
Use key 1 and 2 to change the value. Key 1 is for changing the number
(1-9), key 2 is for changing the position of the cursor. Confirm by pressing
key 3 for >2 sec.
Increase Decrease Left Right
Confirm
The following screen is vissable:
In
Ind 4
you set the
Digital overall filter.
Set the overall filter to effect all indicator
signals used in the device. 0dB means no effect and –50dB is the strongest
damping. Choose between -:
0dB
, 1:
-
6dB
, 2:
-
12dB
, 3:
-
24dB
, 4:
-
30dB
,
5:
-
36dB
, 6:
-
42dB
and 7:
–50dB
.
41