SGM700 Digitizer
First use of indicator -continue-
Calibration Settings -continue-
The indicator now shows CP2 to calibrate the
gain point (CP2)
.
And will automatically jump to:
Use key 1 and key 2 to enter the reference value. Key 1 is used for changing the
number (1-9), key 2 is used for changing the position of the cursor. Load the
weigher with the reference value and press key 3 >2 sec.
Up
Down Left Right
Confirm
Done succesfully the following screen is visible:
Press key 3 <2 sec. to go back to
Configuration Menu.
Press key 3 <2 sec. to go back to
main weigher display.