7
Sampling System Junction Box
Main Electronics and Control Assemblies
Termination / User
connection Box
Dry Sample in
¼ Tube Fitting
vent
Air Intake
PT605 Analyzer – Block Diagram
OPERATION
Overview
The following figure illustrates an overview of the units operation. Refer to figure 1 for
complete system layout.
When the power is applied to the unit it will display startup messages such as current alarm set
points, background or interference gas settings and sensor calibration data.
After the startup messages the system will begin the warm up cycle for approximately 10
minutes (600 seconds). This allows the sampling system to clear and perform initial diagnostic
tests.
H
2
S Measurement:
For the measurement of H2S concentration the analyzer performs sample &
purge routine. The analyzer samples the on-stream gas for 90 seconds and then purges for 180
seconds. At the end of sample cycle the internal solenoid valve is switched off and ambient air
is pumped into the sampling system to purge. The final H2S reading is updated on the display,
4-20mA output is generated and alarm relays are energized if the H2S concentration exceeds
the preset alarm levels.
Sample Conditioning
Pressure Regulators
External Filters
CO2
Sensor Asmbly
Main CPU Board Assembly
Control Logic
Flow Meter +
Coalescing Filter
H2S
Sensor Asmbly
Solenoid
Valve
Gas Sample
routing & Vent
block with tube
Purging Pump
Terminal Blocks
4-20mA
Alarm Relays
AC / DC power Input
Sensor Signal conditioning
AC-DC , DC-DC power convertors. Alarm
relays