STEP 4.
After purge cycle the following message will be displayed.
The unit is now ready for gas standard. Apply the CO2 calibration gas and make sure the
flow meter is set at 1.0 SCFH. Calibration gas must be applied for at least 2 minutes. When the
Sensor Signal reading has stabilized then by using ▲ and ▼ arrow keys adjust and match the
CO2 concentration on the LCD with the CO2 concentration of the calibration gas being applied.
Press ENTER key to accept the change.
STEP 5.
Press the RESET switch when finished to save the calibration data and return to
normal operation.
If calibration is not successful the error message is displayed and the analyzer defaults back to
the previous calibration data.
Apply Calib Gas
CO2 = 0.0
SensrSignl = 0.4 v