33
Modbus Registers
Description
40011
H2S Calibration Data: GFactor
40012
H2S Calibration gas used
40013
H2S Calibration data: Zero Set point
(integer value of volts /100)
For example 90 means 0.90 volts of zero or baseline
sensor signal voltage
40014
H2S Calibration data: Span Set point
(integer value of volts /100)
For example 160 means 1.60 volts of net span
sensor signal
40015
CO2 Calibration gas used
40016
CO2 Calibration data: Zero Set point
(integer value of volts /100)
For example 90 means 0.90 volts of zero or baseline
sensor signal voltage
40017
CO2 Calibration data: Span Set point
(integer value of volts /100)
For example 160 means 1.60 volts of net span
sensor signal
40018
Malfunction or fault code if any
0
No fault
1
Sensor Drift out of tolerance
2
Unit is in exteded clear cycle. This happens for high
concentration of H2S or an additiona time is needed
to clear the remaining sample gas from the sampling
system.
3
Sensor in fault
40019
Bit map of currently active alram relays.
Bit 0: Set if alarm relay 1 is active, clear otherwise
Bit 1: Set if alarm relay 2 is active, clear otherwise
Bit 2: Set if alarm relay 3 is active, clear otherwise
Bit 3: Set if analyzer in Fault state, clear otherwise
Bit 4 thru 15 are not currently used.