When arrow is pointing to
L
(Latching) or
NL
(non-latching) field use the UP and
DOWN arrow keys to toggle between L and NL and press ENTER when done.
After modifying all 3 fields for an alarm the arrow will jump to next line or next alarm.
Modify both Lo and Hi alarms in similar manner using the UP and DOWN arrow keys
to change the value and ENTER key to accept the change and skip to next data field.
Press the RESET key to exit and save alarm levels
Note:
If an alarm is configured as Latching alarm, and once activated it can only be deactivated by pressing the
RESET switch. Non-Latching alarms are, however, deactivated when the concentration goes below the
alarm set point for that channel.