De-ionizer cartridge Replacement
DI cartridge requires an annual change, although it should also be changed if the
water supply is contaminated with water of a conductivity above the specified limit.
Replacement of the DI column (the column in the middle of the water tank) is by
rotating 90° anti-clockwise then pulling up and out of the tank. The new cartridge is
put in by lowering into the recess and twisting 90° clockwise to lock in position.
1. Depressurise the generator and disconnect power.
To replace the DI cartridge, lift
the hinged lid back, then remove the de-ionizer column by turning 90° anticlockwise
and pulling upwards. .
2. Drain the tank using the drain port on the rear of the generator. Once tank is empty,
wipe down with a clean, lint-free cloth.
3. Lower the new DI cartridge into the recess and rotate 90° clockwise to lock in
position, then close the lid.
4. The unit can now be re-filled with water and switched on.