Page 13
Attention - Important Notice to All Users
These instructions must be read thoroughly, section by section and
understood before installation and operation of your Peak Scientific
Hydrogen Generator. Please follow the appropriate safety standards
for handling hydrogen gas and other gases as determined by the laws
and regulations of your country.
Generator is for indoor use only.
Please save the product packaging for storage or future shipment of the generator.
1. For your own personal safety and to prevent accidental damage to your generator
please ensure that you read and understand our User Manual.
2. Do not store away the generator when not in operation. External water bottles must
be made of a dark coloured glass or Nalgene in order to prevent biological growth
(available from Peak Scientific). It is recommended that you check the water quality
weekly to prevent biological growth developing in the tank and tubing. Failure to
do this will greatly reduce the life of the generator and render the 2 year generator
warranty void.
3. If you are using a manual fill system, the tank filter must be replaced after filling the
tank with water.
4. If your system requires an external tank, do not place the external water bottle above
the height of the generator. The bottle may be placed at the same level or below the
generator.
5. Use ASTM Type I or Type II water only. (<1
µ
S conductivity / >1M
Ω
resistivity). Failure
to use the correct quality water will invalidate the cell and system warranty.
6. Failure to maintain the de-ionizer cartridge as recommended will reduce the
operational life of your generator and render its 2 year generator warranty void.
7. Do not attach any additional tanks or gas cylinders to your generator without first
consulting Peak Scientific.
8. Ensure your entire installation, both pipe work and analysers, are completely gas
tight. Even minor leaks will affect the operation of your generator and can reduce
its operational life.
9. Check that the output capacity of your generator meets the total gas requirements
of your application. If in doubt, consult your dealer or Peak Scientific.
10. Ensure the ambient temperature inside your laboratory does not rise above
35°C/95°F. Extremes of operating temperature will reduce cell life.
WARNING
WARNING