Page 22
Installation
Location
The generator should be located on a flat, level surface for operation - this will ensure
that the water level sensors within the unit operate correctly. Some components in the
generator produce heat during operation and therefore an adequate airflow must be
maintained around the generator to allow the ventilation system to perform efficiently.
If the generator is stored in an enclosed space the environment must be controlled via
an air conditioner or extraction fan. The provision must be made to allow the volume
of air in the room to be changed 5 times per hour. The risk with hydrogen is a leak into
the environment raising the hydrogen content to the explosive limit. The rear of the
generator will become warm to the touch during operation - a minimum clearance of 15
cm (6”) from other bodies is recommended.
Do not obstruct or connect the vents to any application. Failing to observe this could
lead to permanent damage of the generator.
Do not expose the unit to freezing conditions. Please adhere to the operating
temperatures below:
Do not position the equipment so that it is difficult to disconnect the mains plug.
Minimum Operating Ambient Temperature:
10 °C (50 °F)
Maximum Operating Ambient Temperature:
35 °C (95 °F)