Page 27
Automatic Filling
(Automatic Fill Upgrade units only)
The Precision Hydrogen generator uses an automatic water filling system as an optional
upgrade, whereby sensors in the unit’s internal water tank detect when the water level
reaches a minimum.
This then refills the tank using a rear connection to an external DI water supply.
The unit should be connected to the water supply as shown below.
Using the length of FEP tubing and the hose barb fitting from the fittings kit, attach the
barb to the tubing. Then connect one end to the water supply and the other to the Fill
port at the rear of the Precision Hydrogen.
For efficient water filling the external supply should be no more than 2 metres away
from the connection on the generator.
The DI Water Supply source MUST NOT be located higher than
the height of the surface that the Precision Hydrogen generator is
placed upon. A supply located higher than specified will result in the
generator’s internal water bottle over filling.
A pressurized DI Water Supply Source may only be used if the
Pressurised Water Feed optional upgrade is purchased.
WARNING