Page 16
Technical Specification
Precision Hydrogen SL
Environment
100 cc
200 cc
Minimum Operating Ambient Temperature
10°C (50°F)
Maximum Operating Ambient Temperature
35°C (95°F)
Maximum Altitude
3000 m
Maximum Relative Humidity
90% Non-Condensing
Minimum Storage Temperature*
10°C (50°F)
Maximum Storage Temperature*
25°C (77°F)
Generator Outlets
Max Flow Rate
100 cc/min
200 cc/min
Purity
99.9995%
Gas outlet
1 x 1/8” Swagelok compression fitting
Delivery pressure
100 psi (6.9 bar)
Electrical Requirements
Voltage
100-/-10%
Frequency
50/60 Hz
Current
6A Max.
Input Connection
C14 Connection
Power Cord (Supplied)
C13 socket to local connection (10A minimum)
Pollution Degree
2
Installation / Over Voltage Category
II
General
Generator Dimensions cm
16.0 (6.3”) x 38.5 (15.2”) x 25.0 (9.8”)
Generator Weight
8 kg (17.6 lbs)
Shipping Crate Dimensions cm
23.4 (9.2”) x 59.4 (23.4”) x 34.2 (13.5”)
Shipping Weight
11 kg (24.2 lbs)
Noise Level @ 1 m
<55 dB
Heat Output
983 BTU / Hr
Water Purity Requirement
ASTM Type II (<1 µS/cm / >1 MΩ-cm)
* When taken out of storage the Generator should be allowed to acclimatize at room
temperature for a minimum of 3 hours before operation.