Page 13
Technical Specification
Genius 1052
Environment
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
35°C (95°F)
Maximum Altitude
3000 m
Maximum Relative Humidity
80% Non-Condensing
Minimum Storage Temperature*
-20°C (-4°F)
Maximum Storage Temperature*
60°C (140°F)
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature
for a minimum of 3 hours before operation.
Please ensure generator is situated in a well-ventilated environment
Generator Outlets
Maximum Heating Outlet Flow Rate
20L/min (0.703cfm) @ 3.1 bar (45psi)
Maximum Cooling Outlet Flow Rate
12L/min (0.424cfm) @ 2.07 bar (30psi)
Maximum Dewpoint
≤-50°C (-58°F)
N2 Purity
≤5%
Particles
<0.01µm
Phthalates
NONE
Suspended Liquids
NONE
Gas Outlets
2 x ¼” BSPP
Drain outlet
1 x ¼” BSPP
Pressure gauges
1
Electrical Requirements
Voltage
220-240 VAC +/- 10%
Frequency
50/60 Hz
Current
7 Amps
Input Connection
C20 Plug
Power Cord
C19 socket to local connection (10A minimum)
Pollution Degree
2
Installation / overvoltage category
II
General
Dimensions cm (inches) WxDxH
60 x 75 x 71.2 (23.6 x 29.5 x 28)
Generator Weight Kg (lbs)
112 (246)
Shipping Weight Kg (lbs)
137 (301)
Содержание Genius 1050
Страница 34: ...Page 34...
Страница 35: ...Page 35...