Technical Specification
Genius 1051
Environment
Genius 1051
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
35°C (95°F)
Maximum Altitude
3000 m
Maximum Relative Humidity
80% Non-Condensing
Minimum Storage Temperature*
-20°C (-4°F)
Maximum Storage Temperature*
60°C (140°F)
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature
for a minimum of 3 hours before operation.
Please ensure generator is situated in a well-ventilated environment
Generator Outlets
Maximum Gas Output Pressure
6.9 bar (100 psi)
Maximum Nitrogen Outlet Flow Rate with
25 L/min (0.88 cfm) with 0 L/min Dry Air**
Maximum Dry Air Outlet Flow Rate
25 L/min (0.88 cfm) with 0 L/min Nitrogen**
** up to 25LPM of Air & Nitrogen
Maximum Dewpoint
-50°C (-58°F) N2 / -20°C (-4°F) Dry Air
N2 Purity
≥97%
Particles
<0.01µm
Phthalates
NONE
Suspended Liquids
NONE
Gas Outlets
2 x ¼” BSPP
Drain outlet
1 x ¼” BSPP
Pressure gauges
2
Air shut-off switch
Yes
Electrical Requirements
Voltage
220-240 VAC +/-10%
Frequency
50/60 Hz
Current
7 Amps
Input Connection
C20 Plug
Power Cord
C19 socket to local connection (10A minimum)
Pollution Degree
2
Installation / overvoltage category
II
General
Dimensions cm (inches) WxDxH
60 x 75 x 71.2 (23.6 x 29.5 x 28)
Generator Weight Kg (lbs)
115 (253)
Shipping Weight Kg (lbs)
136 (299)