WARNING
Electrical Connection
Connect the generator to an appropriate 230 volt single-phase supply, refer to the
generator serial plate for input specification and ensure your supply matches the
requirements.
If the appropriate power cord is not supplied, a new plug with correct rating can be
fitted by a qualified electrician. If an appropriate mains power cords is not supplied
or a substitute one is used then ensure that all components of it the plug, cord
and connector have adequate ratings for the generator and appropriate approvals
for the country of use. Failure to do so could cause damage to the generator or risk
overloading of the power cord
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check the information on the serial plate. The mains supply must be of
the stated AC voltage and frequency.
EARTH/GROUND (E):-
Green & Yellow or
Green
LIVE (L):-
Brown
or
Black
Neutral (N):-
Blue
or
White
Electrical requirements are 230VAC n/- 10% depending on chosen model.
However, running continuously at voltages outwith this is not recommended. Extended
periods at extremes can have a detrimental effect on the operation and life of the
generator.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
To ensure that the correct voltage is being supplied to the generator, the generator
comes equipped with an inbuilt voltmeter. This is located on the underside of the
compressor compartment, and will measure and display the mains voltage that is being
supplied to the generator.
The voltmeter should be checked prior to the initial purge of the system.
If the reading is 195V to 219V, then we would highly recommend fitting a
Dual Tap
Transformer 06-3200
. This can be ordered directly from Peak Scientific.