Page 25
Genius 1051 Gas Flow Combinations
The Genius 1051 are configured in such a way that it is possible to change the flow ratio
of nitrogen and dry air.
The system is capable of providing a variety of nitrogen and dry air combinations, the
table below illustrates some typical flow combinations.
Typical Flow
Dry Air (Lpm)
N2 (Lpm)
Combined (Lpm)
1
0
25
25
2
5
20
25
3
10
15
25
4
15
10
25
5
20
5
25
The figures shown above should be viewed as an approximation, in that, so long as the
combined sum results in a total of 25Lpm, the combinations are numerous.
The gas flow ratios are automatically controlled by the demand requirements of the
instrument being supplied.
Air Selector Switch
The Genius 1051 generators come fitted with an Air Selector Switch on the front panel
of the unit. If the switch is in the OFF position, this will produce a nitrogen supply
only. Turning the switch to the ON position will result in a combination of dry air and
nitrogen, at levels specified by the customer’s instrument.