PDi Communication Systems, Inc.   40 Greenwood Lane   Springboro, Ohio 45066 USA   PH 1-800-628-9870  FX 937-743-5664

 

 

Pillow speaker 

A ¼” stereo-style pillow speaker (pendant control) jack is located on the TV’s connector panel on the backside.  This TV is designed to 
work with either a digital pillow speaker that generates digital-style control codes or a single-button analog (switch-style) pillow speaker.  
A rear panel mounted switch allows use of different brands of pillow speakers and supports the major brands: Zenith, Philips, and RCA.   

 

Pillow Speaker and Antenna Connection 

 

 

 

Across-Room Wiring 

 

NOTE: 

A Jumper Cable is NOT supplied with the TV and must be ordered separately.  

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

CAUTION:

 Confirm the pillow speaker type and set the side switch appropriately BEFORE 

connecting the pillow speaker to the TV. Failure to follow this procedure could result in damage to 

the pillow speaker and/or TV. 

SPEAKER 

 

DATA / SWITCH 

COMMON 

SPEAKER 

DATA / SWITCH 

COMMON 

¼” to 6-Pin Jumper Cable 

PD106-416  

¼” to ¼” Jumper Cable 

PD106-417 

ACROSS ROOM WIRING 

 

WARNING:

  DO NOT connect 

the pillow speaker circuit 
Common to earth ground.  
Grounding of the Common will 
defeat the isolation circuitry of 
the TV and possibly expose 
the patient to harmful shock 
currents should a wiring fault 
occur.

 

 

 

 

CAUTION:

  

Some hospital 

cable systems provide 
power voltage on the coaxial 
cable.  This TV is NOT 
designed to be powered via 
a coaxial cable.  DO NOT 
connect to a powered 
coaxial cable.  Damage will 
result to the TV. 

 

 

OR 

Содержание PDI-P26LCD

Страница 1: ...Document Number PD196 145R3 PDI P26LCD PDI P32LCD PDI P40LCD Healthcare grade Televisions ...

Страница 2: ...cart apparatus combination to avoid injury from tip over MAINTENANCE AND SERVICING Servicing is required when the apparatus has been damaged in any way the power cord or plug is damaged liquid has been spilled objects have fallen into the apparatus the apparatus has been exposed to rain or moisture it does not operate normally or it has been dropped Never remove the back cover of the TV this can e...

Страница 3: ...their respective holders PDI and Better Solutions Are Within Reach are registered trademarks of PDI Communication Systems Inc Springboro Ohio LOCATION GUIDELINES The model PDI P26 P32 P40LCD Hospital Grade LCD TV is a specialized LCD TV It is intended for entertainment and educational purposes in a hospital a nursing home medical care center or similar health care facility in which installation is...

Страница 4: ...he pivot pins are retained in the slots 5 Secure with two 10x32 Pilfer screws PDIPPBHSCS103250 6 The TV s tilt can be adjusted by loosening both Pilfer screws adjusting tilt and then tightening the screws Mounting Pieces Wall Mount LCD Mount Purpose PD168 103 PD181 597 For mounting PDI P26LCD TV to wall without DVD and or IPT Modules 1 PD168 203 PD181 674 For mounting PDI P26LCD TV to wall with DV...

Страница 5: ...the clevis to the wall Place the other supplied flat washer on another screw and insert it in the bottom mounting hole Secure it to the wall and tighten both screws 3 Refer to Figure 2 Attach the LCD mount to adapter plate using the four M4 screws PDIPPHMSM4X76 and four 10 Internal Tooth Lock washers PDIITLW10 NOTE The side with the four hex sides will go toward the back of the set 4 Position the ...

Страница 6: ...ion Across Room Wiring NOTE A Jumper Cable is NOT supplied with the TV and must be ordered separately CAUTION Confirm the pillow speaker type and set the side switch appropriately BEFORE connecting the pillow speaker to the TV Failure to follow this procedure could result in damage to the pillow speaker and or TV SPEAKER DATA SWITCH COMMON SPEAKER DATA SWITCH COMMON to 6 Pin Jumper Cable PD106 416...

Страница 7: ...available depending upon the healthcare facility s signal style Selection of the correct signal type is required for the TV to recognize all possible channels and before any channel programming can begin 1 Press SETUP 2 Press CH or CH to highlight Channels 3 Press VOL to enter the Channels menu 4 Press CH or CH to highlight Signal 5 Press VOL or VOL to select Air Cable STD Cable IRC or Cable HRC N...

Страница 8: ...program additional digital channels press CH or CH to highlight Add Digital Signal Press VOL or VOL to select Air NOTE This feature is only available if Mode is set to Digital Only or Analog and Digital 7 Press CH or CH to highlight the Service Level Free Basic or Premium you wish to program The menu displays the current programming status of each level as either Programmed or Blank NOTE A program...

Страница 9: ...ing on Once the TV is on the volume can be set to any allowable level In Sound menu press CH or CH to highlight Power on Volume 1 Press VOL or VOL to select the preferred volume when the TV turns on NOTE Setting the level to LAST causes the TV to remember the last volume level used by the TV prior to turning off 2 Press the SETUP button to return to the preceding menu Cloning Cloning allows quick ...

Страница 10: ...A confirmation screen will appear Choose Yes to begin upload 7 A status line indicates Successful Saving of TV information to the USB Disk 8 Uploading will take a short time Do not eject USB while uploading 9 TV will restart upon completion and Firmware Upgrade is done will appear 10 Remove the USB Disk from the TV Restore Setting to TV 1 Turn the TV On 2 Insert the thumb drive 3 Press CH or CH to...

Отзывы: