© PCE Instruments
28
DC Current
Range
Resolution
Accuracy
Protection against
overcharge
600.0 µA
0.1 µA
± (0.9 % r 5
digits)
Quick fuse 800
mA/1000 V
6000 µA
1 µA
60.00 mA
0.01 mA
600.0 mA
0.1 mA
± (0.9 % r 8
digits)
10.00 A
0.01 A
±(1.5 % r 8
digits)
Quick fuse 10
A/1000 V
AC TRMS Current
Range
Resolution
Accuracy(*)(50 Hz
… 1 kHz)
Protection against
overcharge
600.0 µA
0.1 µA
±(1.2 % r 5
digits)
Quick fuse 800
mA/1000 V
6000 µA
1 µA
60.00 mA
0.01 mA
600.0 mA
0.1 mA
10.00 A
0.01 A
±(1.5 % r 5
digits)
Quick fuse 10
A/1000 V
(*) Accuracy specified from 5 to 100 % of the measuring range, sine wave.
Accuracy PEAK function: ±10 % rdg , AC+DC TRMS Current: accuracy (50 Hz
… 1 kHz): ±(3.0
% r 20 dgt)
Flexible coil Current
Range
Resolution
(50 Hz
… 60
Hz)
(61 Hz
… 1
kHz)
Protection
against
overcharge
30.00 A
0.01 A
±(0.8 % reading
+ 5 dgt)
±(2.4 % reading
+ 5 dgt)
1000 VDC/AC
rms
300.0 A
0.1 A
3000 A
1 A
Diode test
Function
Test current
Max voltage with open
circuit
<1.5 mA
3.3 VDC
Resistance and Continuity test
Range
Resolution
Accuracy
Buzzer
Protection
against
overcharge
600.0
Ω
0.1
Ω
±(0.5 % reading
+ 10 dgt)
>50
Ω
1000 VDC/AC
rms
6.000
kΩ
0.001
kΩ
±(0.5 % reading
+ 5 dgt)
60.00
kΩ
0.01
kΩ
600.0
kΩ
0.1
kΩ
6.000
MΩ
0.001
MΩ
60.00
MΩ
0.01
MΩ
±(2.5 % reading
+ 10 dgt)