© PCE Instruments
27
Thermal imager
Field of view (FOV)/minimum focus distance
21° x 21°/ 0.5 m
Spatial resolution (IFOV)
4.53 mrad
IR resolution
80 × 80 pixels
Thermal sensitivity/NETD
<0.1°C at +30 °C (+86 °F) / 100 mK
Image frequency
50 Hz
Focus mode
Focus free
Focal length
7.5 mm
Focal Plane Array (FPA)/Spectral range
Uncooled microbolometer / 8
–14 μm
Object temperature range
–20 … +260 °C (–4 … +500 °F)
Accuracy
±3 °C (±5.4 °F) or ±3 % of reading
(Environment temperature
10
… 35 °C, object temperature >0 °C)
Accuracy calculated as [% r (num. digits*resolution)] at 18 °C
… 28 °C <75 % RH
DC voltage
Range
Resolution
Accuracy
Input
impedance
Protection
against
overcharge
600.0 mV
0.1 mV
± (0.09 %
r 5
digits)
>10
MΩ
1000 VDC/AC
rms
6.000 V
0.001 V
60.00 V
0.01 V
600.0 V
0.1 V
±(0.2 % reading
+ 5 digits)
1000 V
1 V
AC TRMS Voltage
Range
Resolution
Accuracy(*)
Protection
against
overcharge
(50 Hz
… 60
Hz)
(61 Hz
… 1
kHz)
6.000 V
0.001 V
±(0.8 % reading
+ 5 dgt)
±(2.4 % reading
+ 5 dgt)
1000 VDC/AC
rms
60.00 V
0.01 V
600.0 V
0.1 V
1000 V
1 V
(*) Accuracy specified from 10 to 100 % of the measuring range, sine wave.
Input impedance: >9
MΩ;
Accuracy PEAK function: ±10 % rdg, PEAK response time: 1 ms
AC+ DC TRMS Voltage
Range
Resolution
Accuracy (50
Hz
… 1 kHz)
Input
impedance
Protection
against
overcharge
6.000 V
0.001 V
±(2.4 % reading
+20 dgt)
>10
MΩ
1000 VDC/AC
rms
60.00 V
0.01 V
600.0 V
0.1 V
1000 V
1 V