443B101/443B102
8
Manual 19854 Rev. D
ECN 50523
The Format of a NAK Message
Reason
Byte
Description
‘C’
Check sum error in message received from external
host.
‘D’
Data overflow error
–
the size of the data field was
larger than the message buffer. The buffer is 95
bytes long.
‘F’
A framing error occurred
–
ETX was received before
it was expected.
‘I’
Check sum error occurred while communicating
with internal module.
‘T’
Time-out
–
message not sent, rack/slot address
may be incorrect.
NAK Reason Bytes
Command format is shown below:
Module Type
Three ASCII printable characters. This field identifies the type of module the command
is meant for. For the 443B series, this will be either C01 (for 443B101) or C02 (for
443B102).
Module
Command
Four printable ASCII characters. This field identifies the command for the module.
Data Field
Contains any data associated with the command.
Start
Byte
(0x02)
NAK
(0x15)
Reason
Byte
End
Byte
(0x030
Checksum
Module
Type
Module
Command
Data Field
Содержание 441A101
Страница 14: ...CHASSIS AND POWER SUPPLY 4 Manual 19680 Rev C ECN 50523 Appendix A Figures...
Страница 15: ...CHASSIS AND POWER SUPPLY 5 Manual 19680 Rev C ECN 50523...
Страница 63: ...443B101 443B102 43 Manual 19854 Rev D ECN 50523 Appendix B Figures...
Страница 64: ...443B101 443B102 44 Manual 19854 Rev D ECN 50523...
Страница 65: ...443B101 443B102 45 Manual 19854 Rev D ECN 50523...
Страница 66: ...443B101 443B102 46 Manual 19854 Rev D ECN 50523...
Страница 67: ...443B101 443B102 47 Manual 19854 Rev D ECN 50523...
Страница 68: ...443B101 443B102 48 Manual 19854 Rev D ECN 50523...
Страница 75: ......
Страница 81: ......
Страница 82: ......