Command: TEDR
Purpose:
Returns data stored in the TEDS sensor.
Note: For sensors progr
ammed with “Accelerometer, transfer function v0.91”
(UTID 116225), use MTED in conjunction with TEDR to retrieve all of the stored
data.
Response:
The module responds with a string of data delimited with semicolons.
Command String:
XYCOZTEDR
Description:
X
—
Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y
—
Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z
—
Module type for the target. Specifying "C01" in this argument addresses
a
443B101, specifying “C02” addresses a 443B102.
TEDR
—
Commands the module to return data from the TEDS sensor.
Response String:
Manufacturer Model Number; Serial Number; Sensitivity; Reference Frequency;
Calibration Date; Low Frequency Response; Output Phase;
Sensitivity Direction; Measurement ID; User Data;
Example:
Command String:
“04C01TEDR”
Return String:
PCB 333M07; SN 17704; 100.2 mV/g; F ref 99.6; cal’d
3/21/2001; F hp 0.025 Hz; phase 0; sens dir N/A;
meas ID 0; test sample 4;
The 443B101 in rack 0 and slot 4 is commanded to return the data stored in the
TEDS sensor connected to its input.
Содержание 441A101
Страница 14: ...CHASSIS AND POWER SUPPLY 4 Manual 19680 Rev C ECN 50523 Appendix A Figures...
Страница 15: ...CHASSIS AND POWER SUPPLY 5 Manual 19680 Rev C ECN 50523...
Страница 63: ...443B101 443B102 43 Manual 19854 Rev D ECN 50523 Appendix B Figures...
Страница 64: ...443B101 443B102 44 Manual 19854 Rev D ECN 50523...
Страница 65: ...443B101 443B102 45 Manual 19854 Rev D ECN 50523...
Страница 66: ...443B101 443B102 46 Manual 19854 Rev D ECN 50523...
Страница 67: ...443B101 443B102 47 Manual 19854 Rev D ECN 50523...
Страница 68: ...443B101 443B102 48 Manual 19854 Rev D ECN 50523...
Страница 75: ......
Страница 81: ......
Страница 82: ......