PCB’s  High  Sensitivity  DC  accelerometers  are  designed 
utilizing  a  MEMS 

(Micro-Electro-Mechanical  Systems)

 

silicon  sensor  approach.  The  micro  machining  design 
process  provides  sensor  design  flexibility,  reliability  & 
stability  over  time.  Compared  to  standard  strain  gages, 
silicon  sensor  accelerometers  have  a  much  higher  gage 
factor.  The  sensor  element  of  the  3741  series  features  a 
proof mass, ringframe, and an attachment system between 
the two. These features are micromachined from the same 
single-crystal  silicon  wafer.  The  sensor  is  connected  as  a 
bridge element in the circuit. The electrical characteristics 
of  one  portion  of  the  bridge  increases  while  the  other 
decreases,  when  exposed  to  acceleration.  The  sensor 
element itself is kept small through the use of cutting edge 
processes  such  as  DRIE  (Deep  Reactive  Ion  Etching). 
Ruggedness  is  enhanced  through  the  use  of  mechanical 
stops  on  the  two  outer  wafers  to  restrict  the  travel  of  the 
proofmass.    Gas  damping  is  used  to  mitigate  high 
frequency inputs. An on-board voltage regulator allows an 
excitation range of +6 to +30 Vdc with less than or equal 
to  5  mA  current  draw.  The  positive  output  signal  line 
increases  with  acceleration  while  the  negative  line 
decreases proportionally. The output lines have a common 
mode voltage of +2.5 Vdc above circuit ground. 
 
 
Consult the individual accelerometer data sheet on how to 
order  the  particular  model  most  appropriate  for  your 
application.  Adhesive  mounting  plates  and  triaxial 
mounting  blocks  are  available.  If  you  have any questions 
or concerns regarding optional configurations, consult the 
Vibration  Division’s  product  catalog  or  contact  a  PCB 
factory representative. 
 
 
4.0 INSTALLATION OVERVIEW 
 
PCB’s  high  sensitivity  DC  accelerometers  are  supplied 
with  a  silicone  jacketed,  shielded,  multi-conductor, 
integral  cable  assembly.  With  each  conductor  being  30 
AWG,  Teflon  coated.  The  supplied  cable  length  for  a 
standard  DC  accelerometer  is  10  ft.  or  120  in.  (~3  m). 
Custom  cable  length  requirements  should  be  specified  at 
time  of  ordering.  When  applications  involve  very  long 
cable  runs,  the  user  is  advised  to  take  into  consideration 
the  added  resistance  of  the  cable  wires  and  additional 
capacitance values, as this may affect the voltage available 
at  the  accelerometers  input  voltage  regulator.  Ensure  the 
recommended  torque  is  used  for  the  supplied  4-40 
mounting  cap  screws.  The  3741  is  packaged  in  a  hard 
anodized  aluminum  case, 0.3 in. high, with a footprint of 
0.85 in. by 1.0 in. 
 

 

The  high  sensitivity  DC  accelerometer  output  signal  is 
considered a differential signal which should be routed to 
a differential amplifier or bridge conditioner (see Fig. 6). 
The  bridge  conditioner  should  supply  a  regulated  dc 
supply and have proper amplification features. Although a 

simple  unregulated  DC  voltage  supply  is  an  acceptable 
condition  for  operation,  bridge  conditioning  equipment 
typically  has  ZMO  adjustment  capability  and  standard 
connector  accessories  included  which  provide  a  “plug  & 
play” environment. 
 
Excitation input and Voltage output wiring conform to the 
following color coding practice. 
 
+   Excitation input (red)   
–  excitation input (black), 
+   signal output (yellow)  
–   signal output (white). 
 
 
Multi-pin  connectors:  For  accelerometer  versions  with 
detachable  cabling,  make  connection  by  inserting  the 
sensor’s  mating  pins  onto  the  cable  connector’s  mating 
sockets.  Then thread the connector into place by turning 
the cable connector’s outer shell onto the accelerometer’s 
electrical connector. 
 
4.1 SCREW MOUNT 
This mounting technique requires smooth, flat contact  
surfaces  for  proper  operation  and  is  recommended  for 
permanent  and/or  secure  installations.  Screw  mounting  is 
recommended when testing at high frequencies.  
 
 
STEP  1:    First,  prepare  a  smooth,  flat  mounting  surface, 
then  drill  and  tap  mounting  holes  for  mounting  as  shown 
in  Figure  3  and  in  accordance  with  the  enclosed 
Installation Drawing
 
 
 
 
 

 
Figure 3. Mounting Surface Preparation 
 
 
 
A  precision-machined  mounting  surface  with  a  minimum 
finish  of  32 

µ

in  (0.0008  mm)  is  recommended.    (If,  and 

only  if,  it  is  not  possible  to  properly  prepare  the  test 

Содержание 3741D4HB200G

Страница 1: ...ROMETER Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail...

Страница 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Страница 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Страница 4: ...ion The sensor element is electrically connected in a Wheatstone bridge circuit The integral amplifier provides a low impedance output to follow on instrumentation These particular accelerometers prov...

Страница 5: ...s 10 ft or 120 in 3 m Custom cable length requirements should be specified at time of ordering When applications involve very long cable runs the user is advised to take into consideration the added r...

Страница 6: ...Mounting Base This method involves attaching a base to the test structure then securing the sensor to the base This allows for easy removal of the accelerometer Contact the PCB factory representative...

Страница 7: ...es to reduce the effects of EMI and RFI that is present near electrical motors and machinery Teflon jacketed cabling may be necessary to withstand corrosive environments and higher temperatures Feel f...

Страница 8: ...its provide the capability for continuous monitoring Note High sensitivity DC Accelerometers feature on board differential amplification 7 0 OPERATING NOTE Always operate the accelerometer within the...

Страница 9: ...s and costly errors Therefore in an effort to prevent the common mistakes associated with customer performed calibration this document includes a broad overview of the Back to Back Calibration techniq...

Страница 10: ...usly calibrated over the frequency and or amplitude range of interest 2 the shaker should be selected to provide minimal transverse lateral motion and minimal distortion and 3 the quality of the meter...

Страница 11: ...SERVICE See the supplement sheet contained in this manual for information on our warranty service repair and return policies and instructions When unexpected measurement problems arise call our 24 ho...

Страница 12: ......

Страница 13: ......

Страница 14: ......

Отзывы: