Understand Back-to-Back Calibration limitations.  Do not 
expect the uncertainty of calibration to be any better than 

±

2%.  (In fact, the uncertainty may be as high as 

±

3% or 

±

4%  for  frequencies  <10  Hz  or  >2  kHz.)    Since  large 

sensors  may  affect  high-frequency  accuracy,  verify  that 
the test sensor does not mass load the reference standard.  
Validate 

your 

calibration 

system 

with 

another 

accelerometer  prior  to  each  calibration  session.    Check 
with the manufacturer for exact system specifications. 
 
 
 
 
 
8.5 CONCLUSIONS 
 
Without  an  adequate  understanding  of  dynamics, 
determining  what,  when,  and  how  to  test  a  sensor  is  a 
difficult  task.    Therefore,  each  user  must  weigh  the  cost, 

time,  and  risk  associated  with  self-calibration  versus 
utilizing the services of an accredited laboratory. 
 
 
 
 
9.0 SERVICE 
 
See  the  supplement  sheet,  contained  in  this  manual,  for 
information  on  our  warranty,  service,  repair,  and  return 
policies and instructions. 
 
When  unexpected  measurement  problems  arise,  call  our 
24-hour SensorLine

SM

 to discuss your immediate dynamic 

instrumentation needs with a factory representative. 
Dial 716-684-0001. 
 
 
 

 
 
 
 
 
 
 
 
 
 
  

 

 

 
 
 
 
 
 
 
 

 

 

 
 

3425 Walden Avenue, Depew, NY 14043-2495 USA    Vibration Division toll-free 888-684-0013 

24-hour SensorLine

SM

 716-684-0001    FAX 716-685-3886    E-mail [email protected]    Website www.pcb.com 

 

A PCB GROUP COMPANY                    ISO 9001 CERTIFIED                   A2LA ACCREDITED to ISO 17025 

 

© 2002 PCB Group, Inc. In the interest of constant product improvement, specifications are subject to change without notice. 

PCB and ICP are registered trademarks of PCB Group, Inc.  SensorLine is a service mark of PCB Group, Inc. 

All other trademarks are properties of their respective owners. 

 

Manual Number:  32258 

VIB-High Sensitivity DC MANUAL-

 

ECN Number: 36609 

Printed in U.S.A

 

Manual Revision:  A 

Содержание 3741D4HB200G

Страница 1: ...ROMETER Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail...

Страница 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Страница 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Страница 4: ...ion The sensor element is electrically connected in a Wheatstone bridge circuit The integral amplifier provides a low impedance output to follow on instrumentation These particular accelerometers prov...

Страница 5: ...s 10 ft or 120 in 3 m Custom cable length requirements should be specified at time of ordering When applications involve very long cable runs the user is advised to take into consideration the added r...

Страница 6: ...Mounting Base This method involves attaching a base to the test structure then securing the sensor to the base This allows for easy removal of the accelerometer Contact the PCB factory representative...

Страница 7: ...es to reduce the effects of EMI and RFI that is present near electrical motors and machinery Teflon jacketed cabling may be necessary to withstand corrosive environments and higher temperatures Feel f...

Страница 8: ...its provide the capability for continuous monitoring Note High sensitivity DC Accelerometers feature on board differential amplification 7 0 OPERATING NOTE Always operate the accelerometer within the...

Страница 9: ...s and costly errors Therefore in an effort to prevent the common mistakes associated with customer performed calibration this document includes a broad overview of the Back to Back Calibration techniq...

Страница 10: ...usly calibrated over the frequency and or amplitude range of interest 2 the shaker should be selected to provide minimal transverse lateral motion and minimal distortion and 3 the quality of the meter...

Страница 11: ...SERVICE See the supplement sheet contained in this manual for information on our warranty service repair and return policies and instructions When unexpected measurement problems arise call our 24 ho...

Страница 12: ......

Страница 13: ......

Страница 14: ......

Отзывы: