For  these  reasons,  it  is  recommended  that  a  recalibration 
cycle be established for each accelerometer.  This schedule 
is unique and is based on a variety of factors, such as: extent 
of  use,  environmental  conditions,  accuracy  requirements, 
trend  information  obtained  from  previous  calibration 
records,  contractual  regulations,  frequency  of  “cross-
checking” 

against 

other 

equipment, 

manufacturer 

recommendation,  and  any  risk  associated  with  incorrect 
readings.    International  standards,  such  as  ISO  10012-1, 
provide  insight  and  suggest  methods  for  determining 
recalibration  intervals  for  most  measuring  equipment.  
With  the  above  information  in  mind  and  under  “normal” 
circumstances,  PCB  conservatively  suggests  a  12-  to  24-
month  recalibration  cycle  for  most  piezoresistive 
accelerometers.   
 
NOTE:  It  is  good  measurement  practice  to  verify  the 
performance  of  each  accelerometer  with  a  calibration 
device before and after each measurement. 
 
 
 
8.1 RECALIBRATION SERVICE 
 
PCB  offers  recalibration  services  for  our  piezoresistive 
accelerometers,  as  well  as  units  produced  by  other 
manufacturers.    Our  internal  metrology  laboratory  is 
certified  to  ISO  9001,  accredited  by  A2LA  to  ANSI/IEC 
17025  and  ANSI/NCSL  Z540-1,  complies  with  ISO 
10012-1  (and  former  MIL-STD-45662A),  and  uses 
equipment directly traceable to NIST.  Our investment in 
equipment,  traceability  and  conformance  to  industry 
standards  ensures  accurate  calibration  against  relevant 
specifications, in a timely fashion. 
 
 
 
8.2 BACK-TO-BACK CALIBRATION THEORY 
 
Many  companies  choose  to  purchase  the  equipment 
necessary to perform the recalibration procedure in house.  
While  this  may  result  in  both  a  savings  of  time  and 
money,  it  has  also  been  attributed  to  incorrect  readings 
and  costly  errors.    Therefore,  in  an  effort  to  prevent  the 
common  mistakes  associated  with  customer-performed 
calibration,  this  document  includes  a  broad  overview  of 
the  Back-to-Back  Calibration  technique.    This  technique 
provides  a  quick  and  easy  method  for  determining  the 
sensitivity  of  a  test  accelerometer  over  a  wide  frequency 
range.   
 
Back-to-Back  Calibration  is  perhaps  the  most  common 
method  for  determining  the  sensitivity  of  piezoresistive 
accelerometers.    This  method  relies  on  a  simple 
comparison  to  a  previously  calibrated  accelerometer, 
typically referred to as a reference standard.   
 

 

Figure 9. Reference Standard Accelerometer 
 
These high-accuracy devices, which are directly traceable to 
a recognized standards laboratory, are designed for stability, 
as  well  as  configured  to  accept  a  test  accelerometer.    By 
mounting a test accelerometer to the reference standard and 
then  connecting  this  combination  to  a  suitable  vibration 
source,  it  is  possible  to  vibrate  both  devices  and  compare 
the  data  as  shown  in  Figure  10.    (Test  set-ups  may  be 
automated and vary, depending on the type and number of 
accelerometers being calibrated). 
 

Figure 10. Typical Back-to-Back Calibration System 
 
Because the acceleration is the same on both sensors, the 
ratio  of  their  outputs  (V

T

/V

R

)  must  also  be  the  ratio  of 

their  sensitivities.    With  the  sensitivity  of  the  reference 
standard  (S

R

)  known,  the  exact  sensitivity  of  the  test 

sensor  (S

T

)  is  easily  calculated  by  using  the  following 

equation: 
 

 

S

= S

R

 (V

T

/V

R

 
By varying the frequency of the vibration, the sensor may 
be  calibrated  over  its  entire  operating  frequency  range.  
The  typical  response  of  an  unfiltered  accelerometer  is 
shown in Figure 11. 

 

Figure 11. Typical Test Accelerometer Response 
 
8.3 PCB CALIBRATION PROCEDURE 
Numerous precautions are taken at PCB to insure accurate 
and  repeatable  results.    This  section  provides  a  brief 
overview of the primary areas of concern. 

Содержание 3741D4HB200G

Страница 1: ...ROMETER Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail...

Страница 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Страница 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Страница 4: ...ion The sensor element is electrically connected in a Wheatstone bridge circuit The integral amplifier provides a low impedance output to follow on instrumentation These particular accelerometers prov...

Страница 5: ...s 10 ft or 120 in 3 m Custom cable length requirements should be specified at time of ordering When applications involve very long cable runs the user is advised to take into consideration the added r...

Страница 6: ...Mounting Base This method involves attaching a base to the test structure then securing the sensor to the base This allows for easy removal of the accelerometer Contact the PCB factory representative...

Страница 7: ...es to reduce the effects of EMI and RFI that is present near electrical motors and machinery Teflon jacketed cabling may be necessary to withstand corrosive environments and higher temperatures Feel f...

Страница 8: ...its provide the capability for continuous monitoring Note High sensitivity DC Accelerometers feature on board differential amplification 7 0 OPERATING NOTE Always operate the accelerometer within the...

Страница 9: ...s and costly errors Therefore in an effort to prevent the common mistakes associated with customer performed calibration this document includes a broad overview of the Back to Back Calibration techniq...

Страница 10: ...usly calibrated over the frequency and or amplitude range of interest 2 the shaker should be selected to provide minimal transverse lateral motion and minimal distortion and 3 the quality of the meter...

Страница 11: ...SERVICE See the supplement sheet contained in this manual for information on our warranty service repair and return policies and instructions When unexpected measurement problems arise call our 24 ho...

Страница 12: ......

Страница 13: ......

Страница 14: ......

Отзывы: