Page 36
NodE 2
SET POINT 1 OPERATION MODE
Select 4 subsequent operation criteria of set point 1: [1308]
Compare with net weight, gross, weight, or peak. In the latter case, comparison occurs with the last
acquired peak value, even if the peak function is not enabled.
NET
The relay output is enabled in the Net Weight mode. [0]
GROSS The relay output is enabled in the Gross Weight mode. [1]
Default: GROSS
Select the output status whether normally open or closed: [1309]
n.oPEn
Relay 1 is normally open. [0]
n.CLoSE Relay 1 is normally closed. [1]
Default: n. oPEn.
Select whether positive or negative values should be compared: [1310]
PoSIt..
The output operates with positive weight. [0]
nEGAt.
The output operates with negative weight. [1]
ALL:
The output operates both with positive and with negative weight. [2]
Default: PoSIt
Select whether stable only or both stable and unstable weight values should be compared: [1311]
norMAL
Output 1 is enabled with unstable weight. [0]
StAbLE
Output 1 is enabled with stable weight. [1]
Default: norMAL
HySt. 2
SET POINT 2 HYSTERESIS [1313]
Hysteresis value vs. the set value set point.
Value: 0 to Capacity
Default: 2
SAME PARAMETERS FOR SET POINTS UP TO 6