![Pavone Systems MC 315 Series Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/pavone-systems/mc-315-series/mc-315-series_technical-manual_763999029.webp)
Page 27
AUto-0
AUTO-ZERO UPON SWITCH-ON [1204-1205]
This parameter defines the value of the maximum weight that can be zeroed when the unit is switched
on. Such operation corresponds to zero-point calibration of the system and is only performed if weight
is stable and below the set-up value.
Value:
0 to capacity. Default: 0
0-trAC
ZERO TRACKING* [1206]
Set-up of the zero tracking value in divisions per second. This function consists in executing automated
zero-point calibration when weight changes slowly in time, as controlled by this parameter. Select
NONE to disable this function. The maximum weight resettable from this function is 2% of system ca-
pacity (parameter SYS FS)
Value
Variation
0
Control excluded
1
0.5 div/sec
2
1 div/sec
3
2 div/sec
4
3 div/sec
Default: 0
0-bAnd
ZERO BAND* [1207]
This parameter defines the number of divisions that can be zeroed by pressing the front zero key or
via the associated Input.
Value: 0 to 200.
Default: 100
G-CAL
GRAVITY IN THE PLACE OF CALIBRATION [1209-1210]
Set-up of the gravity value in the place of calibration. (**)
Values: 9.70000 to 9.84000
Default: 0
G-uSE
GRAVITY IN THE PLACE OF USE [1211-1212]
Set-up of the gravity value in the place where the unit is used. (**)
Values: 9.70000 to 9.84000
Default: 0
(*) These parameters are only used in case of FREE operation or in case of METRIC operation with ca-
libration bridge enabled.
(**) Following a change of these values, weight calibration is adjusted according to the entered accele-
ration and gravity parameters. This function is only executed if both parameters, “G-CAL” and “G-USE”,
are set to a value other than 0.