WEIGHING PARAMETERS
F1LtEr
WEIGHT FILTER VALUE [1208]
Low filter values speed up weight updating to the detriment of stability. High filter values slow down
weight updating and make it more stable.
Factor
(Hz)
Settling Time
(mS)
Freq ADC
(Hz)
Nr of
reading
Monotony Time
(mS)
Oscillation time
(mS)
Oscillation
range (div)
F0
100
50
5
20
4000
10
F1
300
50
15
40
3000
12
F2
500
50
25
80
2500
16
F3
700
50
35
100
2000
20
F4
900
50
45
240
1500
25
F5
1100
50
55
240
1500
25
F6
1300
50
65
300
1500
25
F7
1500
50
75
400
1500
25
F8
1700
50
85
500
1200
30
F9
1900
50
95
600
1000
30
Value
Variation
F0
High-speed weight reading, low stability
F9
Low-speed weight reading, high stability
Default: F5
uSE
UNIT OPERATION *
Unit operation mode selection. When shifting from FREE to METRIC operation, setting confirmation
requires the password-based authentication of the authorized staff.
Value
Variation
Free
Free operation. [0]
Trade
METRIC operation. [1]
Default: Free
SyS FS
SYSTEM FULL SCALE * [1201-1202]
Maximum system capacity set-up, expressed in the weight division value.
Value:
0 to Cell capacity.
Default: Cell capacity
NotIon
WEIGHT STABILITY* [1203]
This parameter defines the number of divisions required to consider weight as stable.
A high number of divisions allows the indicator to quickly record weight stability, which is required to
execute the tare and print commands.
Value from 0 to 4
Default: 2