Page 69
The following table lists the registers of the output area (written by the master and acquired by the
instrument), common to all PROFIBUS, PROFINET, ETHERCAT, ETHERNET / IP Fieldbuses.
The registers are 16 bit in size. The registers written by the master in the output area, are read by the
instrument at a fixed frequency of 125 Hz. (80 Hz in case of PROFIBUS)
The size of the output area configured in the master fieldbus must match the size configured in the
instrument.
OUTPUT DATA AREA
Byte
Register
address OUTPUT AREA REGISTER
Note
1-2
0
Command Register
See relevant table.
3-4
1
Data Register (MSB)
INT. value - Most significant word (See Table)
5-6
2
Data Register (LSB)
INT. value - Least significant word (See Table)
7-8
3
Monitor register
This value corresponds to the same register in the input area.
9-10
4
Set-Point 1 (MSB)
INT. value - Most significant word
11-12
5
Set-Point 1 (LSB)
INT. value - Less significant word
13-14
6
Set-Point 2 (MSB)
INT. value - Most significant word
15-16
7
Set-Point 2 (LSB)
INT. value - Less significant word
17-18
8
Capacity of load cells (MSB)
INT. value - Most significant word
19-20
9
Capacity of load cells (LSB)
INT. value - Less significant word
21-22
10
Sensitivity of load cells
INT. value
23-24
11
Weight division value
See relevant table.
25-26
12
Decimals
See relevant table.
27-28
13
Fixed Tare (MSB)
INT. value - Most significant word
29-30
14
Fixed Tare (LSB)
INT. value - Less significant word
31-32
15
Stand-by function
INT. value
33-34
16
Keypad Lock function
See relevant table.
35-36
17
Password function
INT. value
37-38
18
Peak function
See correspondence on page 51
39-40
19
Data-Logger function
See correspondence on page 52
41-42
20
Data-Logger Trigger
See correspondence on page 52
43-44
21
Data-Logger function
See correspondence on page 52
45-46
22
Filter factor
See correspondence on page 49
47-48
23
Output rate ADC
See correspondence on page 49
49-50
24
Number of readings on average INT. value
51-52
25
Monotony Time
INT. value
53-54
26
Oscillations Time
INT. value
55-56
27
Oscillations Range
INT. value
57-58
28
Full Scale (MSB)
INT. value - Most significant word
59-60
29
Full Scale (LSB)
INT. value - Less significant word
61-62
30
Weight stability
See correspondence on page 47
63-64
31
Auto zero when switching on (MSB) INT. value - Most significant word
65-66
32
Auto zero when switching on (LSB) INT. value - Less significant word
67-68
33
Zero tracking
See correspondence on page 48
69-70
34
Resettable Divisions (key >0<)
INT. value
71-72
35
Input 1 function
See correspondence on page 44
73-74
36
Input 2 function
See correspondence on page 44
75-76
37
Input 1 mode —Function
See correspondence on page 44