Page 36
COM 2 PARAMETERS WHEN PRESENT RS485
C2 Nod..
COM2 OUTPUT MODE
Selecting the value transmitted on output RS 485.
Values that can be selected:
nEt
GroSS
PEAk
Default: nEt
C2Prot...
COM2 PROTOCOL
It defines how to use the RS485 serial port:
Values that can be selected:
None:
Serial communication OFF
Contin:
Continuous transmission of the weight string. It can be used, for example, to drive a weight
repeater. See details in the relevant section.
on deM:
When the operator presses the relevant front key or via Input 2, a weight string is
transmitted. The command is accepted if the weight is stable. Between two subsequent transmissions
the weight must undergo a variation of 20 divisions.
Autom:
A weight string is sent automatically when theweight stabilises at a value above the minimum
weight (20 divisions). Variation of the weight between two consecutive transmissions must be at least
20 divisions.
Slave:
ASCII protocol. See details in the relevant section.
Modbus:
MODBUS RTU Protocol See details in the relevant section.
Default: Slave
C2baud.
COM2 BAUD RATE
Defines the baud rate of serial port RS485.
The value must be set at the same value as PC/PLC or remote display.
Values that can be selected:
1200
2400
4800
9600
19200
38400
57600
115200
Default: 9600