P
AGE
72
THERMOCOUPLE
TYPE
RANGE
TYPE
RANGE
J
-130 to 760
°
C
E
0 to 750
°
C
-200 to 1400
°
F
0 to 1400
°
F
K
-130 to 1370
°
C
B
200 to 1800
°
C
-200 to 2500
°
F
400 to 3300
°
F
T
-200 to 400
°
C
N
0 to 1300
°
C
-330 to 750
°
F
0 to 2370
°
F
R
200 to 1650
°
C
C
200 to 2300
°
C
400 to 3000
°
F
390 to 4170
°
F
S
200 to 1650
°
C
400 to 3000
°
F
RTD
VOLTS
MILLIVOLTS
100 OHM
0 to 5VDC
0 to 25mVDC
-140 to 440
°
C
1 to 5VDC
0 to 50mVDC
-220 to 750
°
F
10 to 50mVDC
0 to 25mVDC
MILLIAMPS
4-20mA input is accommodated via
10-50mV or 1-5V with the addition
of the appropriate external shunt resistor
SENSOR FAULT DETECTION
Displays Hi or Lo process input for thermocouple or RTD inputs (10% above or below range)
and sensor break, SnSr. On/Off outputs and events go off, proportional outputs go to 0%
output. Sensor fault detection is not functional for 0 to 5 VDC.
CONTROL ADJUSTMENTS
On/Off Hysteresis
0 to 300 units
Proportional Band
1 to 3000 units
Manual Reset
-1500 to 1500 units
Auto Reset
0.0 to 100 repeats/min.
Rate
0.0 to 10.0 min.
Cycle Time
1 to 240 sec.
Position Proportioning
0.0 to 50.0%
Sensitivity
First Output Position
-1000 to 1000 units
Spread
-1000 to 1000 units
(Second Output Position)