P
AGE
61
Er10
- ADC Reference
1. Perform the CAL2 procedure as described in the
Voltage Error
Calibration section (page 48).
Er11
- Cold Junction
1. Be sure the Cold Junction Sensor is firmly attached
Compensation Error
to terminals 2 and 4.
2. Perform the CAL3 procedure as described in the
Calibration section (page 49).
Er12
- CAL2 Voltage Error
1. Check that 50mVDC is properly connected to the
instrument and is within the tolerance limits as
indicated in the CAL2 procedure of the Calibration
section (page 48).
2. Loosen the front panel screw and remove the
instrument from the housing. Inspect the Processor
board (Appendix A-2, page 64) to insure that the input
conditioning jumper JU1 is in the non-volt position.
3. Perform the CAL2 procedure as described in the
Calibration section (page 48)
Er13
- RTD CAL5 Input Error
1. Check that the resistance device is of the correct value
and properly connected to the instrument and is within
the tolerance limits as indicated in the CAL5 procedure
of the Calibration section (page 50).
2. Loosen the front panel screw and remove the
instrument from the housing. Inspect the Processor
board (Appendix A-2, page 64) to insure that the input
conditioning jumper JU1 is in the non-volt position and
that the Option board jumpers (Appendix A-3, page
65) JU2 and JU3 are in the RTD position.
3. Perform the CAL 5 procedure as described in the
Calibration section (page 50).
Er14
- Cold Junction
1. Be sure the Cold Junction Sensor is firmly attached
Compensation Error
to terminals 2 and 4.
2. Perform the CAL3 procedure as described in the
Calibration section (page 49).
Er15
- Ground Reference
1. Perform the CAL2 procedure as described in the
Tolerance Error
Calibration section (page 48).
Er16
- Program/Tune Mode
1. Record all Program and Tune mode parameters.
Checksum Error
Perform the CAL1 procedure as described in the
Calibration section (page 48). Re-enter the Program
and Tune mode parameters.
Er17
- Calibration Checksum
1. Perform the calibration procedures that are needed for
Error
the input sensors that will be used (page 48).