HandiLaz Operator’s Manual
Page 6-1
Chapter 6
Troubleshooting
Problem
Possible Cause
Try…
HandiLaz will not turn on
• Battery missing
• Battery discharged
• Charger not connected
• Charger not powered up
Check the presence and
connection of the internal
battery pack. Charge the
battery. Make sure the charger
is on and the charger power
cable is plugged in.
HandiLaz display is dark or not on
Contrast incorrectly set.
Turn power off and back on.
Adjust the contrast setting to
achieve the required display
level.
Keypad does not respond
Processing lock-up
Cycle power. Turn unit off.
Wait 10 seconds and turn back
on. Ensure connection to any
PC has followed the
recommendations in the
“CountWin” section.
Pump error reported
Obstruction to flow on inlet
Check for obstruction. Do not
insert anything into the inlet
nozzle as this could damage
the sensitive optics.
Printer does not function
• Printer not turned on
• Incorrect DIP settings on
printer
• Connections faulty
Ensure printer powered up.
Ensure DIP settings are as per
this manual.
TRH probe does not function
• Improper probe
connection
• Incorrect software menu
Ensure the probe is pushed
fully home. Ensure the
measure menu is being used
for TRH readings.
Содержание HandiLaz
Страница 8: ...viii List of Figures This page is intentionally left blank...
Страница 36: ...Page 5 6 HandiLaz Operator s Manual Chapter 5 Using Optional Accessories This page is intentionally left blank...
Страница 38: ...Page 6 2 HandiLaz Operator s Manual Chapter 6 Troubleshooting This page is intentionally left blank...
Страница 40: ...Page 7 2 HandiLaz Operator s Manual Chapter 7 Maintenance This page is intentionally left blank...
Страница 42: ...Page A 2 HandiLaz Operator s Manual Appendix A Options and Accessories This page is intentionally left blank...