SECTION 7 – TROUBLESHOOTING 

 

ALARM CODE DISPLAY 

 
When a flow rate deviation alarm occurs or when an alarm occurs during MFC self-diagnostics, the 
operating mode defaults to the operating mode currently selected at ‘

C-16

’ in the function setup, 

‘Operation selection at alarm occurrence‘.  This holds true for all alarm codes except  ‘

AL71

’. 

 
The upper display alternately indicates the alarm codes shown in the table below and the regular display, 
the instantaneous flow rate display.  Alarm codes are displayed only when the instantaneous flow rate, 
operation mode or totalized flow rate is displayed. 
 

ALARM 

CODE 

ERROR 

POSSIBLE CAUSE 

CORRECTIVE ACTION 

AL01 

Flow rate deviation 
lower limit alarm 

Insufficient alarm judgment delay 
time.  Insufficient power supply 
voltage.  Insufficient inlet pressure, 
etc. or excessive operating 
temperature. 

If none of the possible causes 
resolves the problem, contact 
factory. 

AL02 

Flow rate deviation 
upper limit alarm 

Insufficient alarm judgment delay 
time, valve problem, sensor problem, 
etc. 

If no problem found with 
delay time, contact factory. 

AL71 

Valve overheat 
prevention limit is 
actuated 

During the control or fully-open mode, 
the gas is shut off for more than five 
(5) minutes. 

If the gas is shut off for more 
than five (5) minutes, set the 
set flow rate (

SP

) to zero or 

valve fully-closed. 

AL81 

Sensor error 

Sensor problem, particulate in sensor 
or use of helium or hydrogen gas. 

Turn power off and restore 
power.  If sensor not 
restored, contact factory. 

AL91 

I/O correction data 
error 

Data corrupted due to electrical 
noise. 

Contact factory. 

AL92 

Sensor calibration 
data error 

Data corrupted due to electrical 
noise. 

Contact factory. 

AL93 

User setup data error 

Power loss during data transmission. 

Reset data. 

 

AL71

 (valve overheat prevention limit) is available only for Model MPC20.  Despite the current selection 

at ‘

C-16

’ in the function setup, ‘Operation selection at alarm occurrence‘, the valve drive current is forcibly 

limited.  If this situation occurs for thirty (30) minutes or more, the valve is fully closed. 
 
If 

AL81

 (sensor error) occurs, the flow rate will be uncontrolled.  This occurs even if the ‘

C-16

’ function 

setup ‘0:  Control continued (alarm ignored)’ is selected. 
 
When ‘1:  Move to fully closed’ or ‘2:  Move to fully open’ is selected at ‘

C-16

’ in the function setup, the 

alarm display and the operating mode at alarm occurrence can be maintained even after the cause of 
alarm is removed.  When canceling the alarm, perform the alarm reset operation. 
 

Alarm Reset Operation 

 
When the [

ENT

] key is continuously pressed while the instantaneous flow rate is displayed, the alarm will 

reset after three (3) seconds. 
 

Содержание Porter MPC Series

Страница 1: ...MPC SERIES PANEL MOUNT MASS FLOW CONTROLLER INSTALLATION PROGRAMMING USER S MANUAL FM 1038 Rev 0 4 08...

Страница 2: ...G 9 Location 9 Dimensional Data 9 Mounting Procedure 9 External Electrical Connections 11 Section 5 BASIC OPERATION 13 Switching Displays 13 Setting the Flow Rate 15 Selecting the Operation Mode 17 Se...

Страница 3: ...positive shut off is a system requirement install a separate shut off valve When a separate shut off valve is closed the user must command the MFC control valve fully closed by one of the following me...

Страница 4: ...User s Manual FM 1032 One 1 FEATURES The MPC Series Panel Mount MFC is a mass flow controller featuring high performance in a compact package design Integrating the advanced technologies of an ultra...

Страница 5: ...rt stop reset can be remote controlled by external switching input After a reset by key operation the totalization calculation is started automatically However if a reset contact ON is carried out by...

Страница 6: ...parate shut off valve AUTOMATIC RESET OF TOTALIZATION COUNT AT START OF CONTROL FUNCTION Start of control and totalization count reset can be carried out simultaneously by a single action key operatio...

Страница 7: ...0 C 14 to 140 F Humidity Operating Humidity Range 10 to 90 Relative Humidity non condensing Setpoint Input Keypad Operation or External Setpoint Voltage Input Resolution Refer to Table 1 Setpoint Setp...

Страница 8: ...lectrical connector Note 1 SLPM indicates the volumetric flow corrected to 20 C and 1 atmosphere 14 7 PSIA The reference temperature can also be changed to 0 C 25 C and 35 C The controllable flow rang...

Страница 9: ...also displays the operation mode totalized flow rate value lower 4 digits valve drive output parameter setup values or function setup values OPERATION LAMP L Indicates the totalized flow rate is disp...

Страница 10: ...nnection of a dedicated loader cable available separately to the loader jack on the rear of the MFC enables direct communication with a personal computer IN CONNECTOR 1 8 FNPT inlet connector OUT CONN...

Страница 11: ...NG PROCEDURE PROCESS CONNECTION INSTALLATION While installing the inlet and outlet process connections to the inlet and outlet 1 8 FNPT connectors do not hold the housing of the MPC Series Panel Mount...

Страница 12: ...From the front of the panel insert the MFC through the panel cutout until the MFC s bezel rests against the front of the panel With the mounting bracket screws oriented vertically i e 6 o clock and 12...

Страница 13: ...ng power The event output does not exceed the specified output rating of the MFC When driving a relay use a relay with a built in diode for coil surge suppression ELECTRICAL CONNECTOR SPECIFICATIONS P...

Страница 14: ...12 WIRING TYPICAL WIRING...

Страница 15: ...e function setup For details on function setup method refer to SECTION 6 APPLICATION OPERATION 5 If no setting change is made while the multi setting flow rate is displayed the display will automatica...

Страница 16: ...OW RATE DISPLAY When the DISP key is pressed while the operation mode is displayed the L lamp lights and the totalized flow rate value is indicated on the upper display and lower display For example i...

Страница 17: ...when performing incremental changes to the SP value Multi SP Setpoint Setting Mode Multiple Setpoints Per Function Setup C 04 Up to four 4 SP setpoint values setting flow rate values can be programme...

Страница 18: ...splayed indication is automatically returned to the instantaneous flow rate display PROCEDURE FOR CHANGING FLOW RATE IN ANALOG SETTING To change the SP value using an external setpoint voltage confirm...

Страница 19: ...the DISP key is pressed 3 When the DISP key is pressed while the display is flashing step 2 of Selecting the Operating Mode procedure above the operating mode selection is cancelled 4 When 5 Valve ful...

Страница 20: ...onds or longer press DISP key less than 1 second press DISP key or any key is not pressed for 10 seconds or longer press DISP key 1 second or longer flow rate indication 3 Totalized flow rate Totalize...

Страница 21: ...ow rate display FUNCTION SETUP ITEM LIST Display Function Description Setup Options and Descriptions Factory Setting Remarks C 01 Key lock 0 Unlocked 1 Settings other than flow rate setting are locked...

Страница 22: ...nal switch input 1 function assignment 0 C 11 External switch input 2 function assignment 0 Not used 1 Reset totalization 2 Stop totalization count operation 3 Switching of SP number 4 Switching of fl...

Страница 23: ...pressure adjustment 0 0 0 1 MPa 0 14 5 PSIG 1 0 05 0 15 MPa 7 25 21 8 PSIG 2 0 15 0 25 MPa 21 8 36 3 PSIG 3 0 25 0 35 MPa 36 3 50 8 PSIG 4 0 35 0 45 MPa 50 8 65 3 PSIG 5 0 45 0 5 MPa 65 3 72 5 PSIG 2...

Страница 24: ...ence No Display Parameter Description Factory Setting Setting Range Refer to Function Setup Item Remarks 1 0 r 9 1 Flow rate OK judgment range 2 FS 10 0 5 to 100 FS 10 2 0 HYS 1 Flow rate OK judgment...

Страница 25: ...vent output type assignment in the function setup 7 This can be set only when 1 Function enabled is selected for the C 28 analog optional scaling function in the function setup 8 This can be set only...

Страница 26: ...the gas is shut off for more than five 5 minutes set the set flow rate SP to zero or valve fully closed AL81 Sensor error Sensor problem particulate in sensor or use of helium or hydrogen gas Turn po...

Страница 27: ...nstall an upstream filter Contact factory Flow rate does not stabilize Operational differential pressure range is exceeded Excessive supply pressure fluctuations Pressure regulator compatibility Exces...

Страница 28: ...ter Instrument Division and accompanied by a properly completed Returned Goods Authorization All returns are subject to restocking and possible rework charges to be determined by Porter Instrument Div...

Страница 29: ...ons and calibration the customer will be notified of the costs before proceeding with repair or replacement Repaired equipment will be warranted ninety 90 days from date of delivery to the customer or...

Страница 30: ......

Отзывы: