- 25 -
MISCELLANEOUS ERRORS
ERROR
POSSIBLE CAUSE
CORRECTIVE ACTION
Flow rate display does not reach
zero (0) when actual flow stops.
Display does not read
‘
OFF
’
even if valve is fully closed.
▪
Zero point deviation due to
pressure effect.
▪
Gas type setup is incorrect.
▪
Condensation on sensor.
▪
Particulate in sensor.
▪
Match the inlet pressure setting
(function setup
C-20
’
) with the
actual inlet pressure used or the
PV
forced zero function (function
setup
‘
C-29
’
).
▪
Match the gas type setting
(function setup
‘
C-18
’
) with the
actual gas used.
▪
Install an upstream filter.
▪
Contact factory.
Flow rate does not stabilize.
▪
Operational differential pressure
range is exceeded.
▪
Excessive supply pressure
fluctuations.
▪
Pressure regulator compatibility.
▪
Excessive pressure drop
(excessive supply pressure
fluctuations versus flow
changes).
▪
Reduce the supply pressure.
▪
Install a pressure regulator to
regulate supply pressure.
▪
Adjust the pressure regulator
setting or apply the
PV
filter
(function setup
‘
C-23
’
).
▪
Increase the tubing diameter.
Poor accuracy.
▪
Temperature reference not
identical to reference
flowmeter.
▪
Pressure regulator is vibrating
slightly.
▪
Particulate in sensor.
▪
Match the temperature
reference (function setup
‘
C-
19
’
).
▪
Adjust the pressure regulator
setting.
▪
Contact factory.