Parker Porter MPC Series Скачать руководство пользователя страница 21

 

- 19 - 

FUNCTION SETUP 

 
Follow the procedure below to set the functions such as event output type and external switch input 
assignments: 
 

1.  After introducing power to the MFC, press the [

DISP

] key twice (2x) to display the totalized flow 

rate (‘

L

’ lamp lights). 

2.  Press the [

<

] key for three (3) seconds or longer (the upper display indicates 

0.r

9

 [parameter 

setup mode]). 

3.  Again, press the [

<

] key for three (3) seconds or longer (the upper display indicates the function 

setup item number 

C-01

 [function setup mode]). 

4.  Press either the [

] or [

] key until the desired setup item number is indicated. 

5.  Press the [

ENT

] key (the present setting value indicated on the lower display blinks). 

6.  Press either the [

] or [

] key to select the desired setting value.  Press the [

ENT

] key to store the 

selected value.  If the [

DISP

] key is pressed prior to pressing the [

ENT

] key, the setting value 

remains at the previous value without being updated. 

7.  Repeat steps 4, 5 and 6 for any setup item requiring programming.  Once setup items are 

completed, proceed to step 8. 

8.  Press the [

DISP

] key to return to the instantaneous flow rate display, the display shown after 

introducing power. 

 

Note:  If no operation is performed for one (1) minute after entering the function setup mode, the display 

automatically returns to the instantaneous flow rate display. 

 

FUNCTION SETUP ITEM LIST 

 

Display 

Function Description 

Setup Options and Descriptions 

Factory 

Setting 

Remarks 

C-01 

Key lock 

0: 

Unlocked 

1: 

Settings other than flow rate setting 
are locked 

2: 

All settings locked 

The key lock can be canceled even while it is 
enabled. 

LoC

’ is indicated on the display during the key-

locked setting. 

C-02 

Operating mode 
selection (selection by 
key operation) 

0: 

Operating mode selection by key 
operation is disabled. 

1: 

Operating mode selection by key 
operation is enabled. 

The selection of whether operating mode 
selection by key operation (fully 
closed/control/fully open) is “enabled” or 
“disabled” is available.  Refer to instructions for 
“Selecting the Operating Mode”. 

C-03 

Instantaneous flow rate 
setup method 
(instantaneous 

SP

 setup 

method selection) 

0: 

Digital setup (set by key operation) 

1: 

Analog setup (set by external 
analog voltage input) 

When ‘4:  Switching of flow rate setup method’ is 
selected at the external switch input function 
assignment ‘

C-10

’ or ‘

C-11

’, the switching by 

external switch input takes precedence. 

C-04 

Selection of total 
number of 
instantaneous flow rate 
setups (selection of total 
number of 

SP

 values) 

0: 

Number of 

SP

 values = 1 

(

SP-0 

only) 

1: 

Number of 

SP

 values = 2 

(

SP-0

 and 

SP-1

2: 

Number of 

SP

 values = 3 

(

SP-0

SP-1

 and 

SP-2

3: 

Number of 

SP

 values = 4 

(

SP-0

SP-1

SP-2

 and 

SP-3

 

C-05 

Setpoint input voltage 
range selection, at 
analog setting (analog 

SP

 input range 

selection) 

0: 

0 to 5 Vdc input 

1: 

1 to 5 Vdc input 

 

Содержание Porter MPC Series

Страница 1: ...MPC SERIES PANEL MOUNT MASS FLOW CONTROLLER INSTALLATION PROGRAMMING USER S MANUAL FM 1038 Rev 0 4 08...

Страница 2: ...G 9 Location 9 Dimensional Data 9 Mounting Procedure 9 External Electrical Connections 11 Section 5 BASIC OPERATION 13 Switching Displays 13 Setting the Flow Rate 15 Selecting the Operation Mode 17 Se...

Страница 3: ...positive shut off is a system requirement install a separate shut off valve When a separate shut off valve is closed the user must command the MFC control valve fully closed by one of the following me...

Страница 4: ...User s Manual FM 1032 One 1 FEATURES The MPC Series Panel Mount MFC is a mass flow controller featuring high performance in a compact package design Integrating the advanced technologies of an ultra...

Страница 5: ...rt stop reset can be remote controlled by external switching input After a reset by key operation the totalization calculation is started automatically However if a reset contact ON is carried out by...

Страница 6: ...parate shut off valve AUTOMATIC RESET OF TOTALIZATION COUNT AT START OF CONTROL FUNCTION Start of control and totalization count reset can be carried out simultaneously by a single action key operatio...

Страница 7: ...0 C 14 to 140 F Humidity Operating Humidity Range 10 to 90 Relative Humidity non condensing Setpoint Input Keypad Operation or External Setpoint Voltage Input Resolution Refer to Table 1 Setpoint Setp...

Страница 8: ...lectrical connector Note 1 SLPM indicates the volumetric flow corrected to 20 C and 1 atmosphere 14 7 PSIA The reference temperature can also be changed to 0 C 25 C and 35 C The controllable flow rang...

Страница 9: ...also displays the operation mode totalized flow rate value lower 4 digits valve drive output parameter setup values or function setup values OPERATION LAMP L Indicates the totalized flow rate is disp...

Страница 10: ...nnection of a dedicated loader cable available separately to the loader jack on the rear of the MFC enables direct communication with a personal computer IN CONNECTOR 1 8 FNPT inlet connector OUT CONN...

Страница 11: ...NG PROCEDURE PROCESS CONNECTION INSTALLATION While installing the inlet and outlet process connections to the inlet and outlet 1 8 FNPT connectors do not hold the housing of the MPC Series Panel Mount...

Страница 12: ...From the front of the panel insert the MFC through the panel cutout until the MFC s bezel rests against the front of the panel With the mounting bracket screws oriented vertically i e 6 o clock and 12...

Страница 13: ...ng power The event output does not exceed the specified output rating of the MFC When driving a relay use a relay with a built in diode for coil surge suppression ELECTRICAL CONNECTOR SPECIFICATIONS P...

Страница 14: ...12 WIRING TYPICAL WIRING...

Страница 15: ...e function setup For details on function setup method refer to SECTION 6 APPLICATION OPERATION 5 If no setting change is made while the multi setting flow rate is displayed the display will automatica...

Страница 16: ...OW RATE DISPLAY When the DISP key is pressed while the operation mode is displayed the L lamp lights and the totalized flow rate value is indicated on the upper display and lower display For example i...

Страница 17: ...when performing incremental changes to the SP value Multi SP Setpoint Setting Mode Multiple Setpoints Per Function Setup C 04 Up to four 4 SP setpoint values setting flow rate values can be programme...

Страница 18: ...splayed indication is automatically returned to the instantaneous flow rate display PROCEDURE FOR CHANGING FLOW RATE IN ANALOG SETTING To change the SP value using an external setpoint voltage confirm...

Страница 19: ...the DISP key is pressed 3 When the DISP key is pressed while the display is flashing step 2 of Selecting the Operating Mode procedure above the operating mode selection is cancelled 4 When 5 Valve ful...

Страница 20: ...onds or longer press DISP key less than 1 second press DISP key or any key is not pressed for 10 seconds or longer press DISP key 1 second or longer flow rate indication 3 Totalized flow rate Totalize...

Страница 21: ...ow rate display FUNCTION SETUP ITEM LIST Display Function Description Setup Options and Descriptions Factory Setting Remarks C 01 Key lock 0 Unlocked 1 Settings other than flow rate setting are locked...

Страница 22: ...nal switch input 1 function assignment 0 C 11 External switch input 2 function assignment 0 Not used 1 Reset totalization 2 Stop totalization count operation 3 Switching of SP number 4 Switching of fl...

Страница 23: ...pressure adjustment 0 0 0 1 MPa 0 14 5 PSIG 1 0 05 0 15 MPa 7 25 21 8 PSIG 2 0 15 0 25 MPa 21 8 36 3 PSIG 3 0 25 0 35 MPa 36 3 50 8 PSIG 4 0 35 0 45 MPa 50 8 65 3 PSIG 5 0 45 0 5 MPa 65 3 72 5 PSIG 2...

Страница 24: ...ence No Display Parameter Description Factory Setting Setting Range Refer to Function Setup Item Remarks 1 0 r 9 1 Flow rate OK judgment range 2 FS 10 0 5 to 100 FS 10 2 0 HYS 1 Flow rate OK judgment...

Страница 25: ...vent output type assignment in the function setup 7 This can be set only when 1 Function enabled is selected for the C 28 analog optional scaling function in the function setup 8 This can be set only...

Страница 26: ...the gas is shut off for more than five 5 minutes set the set flow rate SP to zero or valve fully closed AL81 Sensor error Sensor problem particulate in sensor or use of helium or hydrogen gas Turn po...

Страница 27: ...nstall an upstream filter Contact factory Flow rate does not stabilize Operational differential pressure range is exceeded Excessive supply pressure fluctuations Pressure regulator compatibility Exces...

Страница 28: ...ter Instrument Division and accompanied by a properly completed Returned Goods Authorization All returns are subject to restocking and possible rework charges to be determined by Porter Instrument Div...

Страница 29: ...ons and calibration the customer will be notified of the costs before proceeding with repair or replacement Repaired equipment will be warranted ninety 90 days from date of delivery to the customer or...